Abstract
We report on a recently developed novel type of large area (62 mm x 62 mm) position sensitive silicon detector with four-corner readout. It consists of a square-shaped ion-implanted resistive anode framed by additional low-resistivity strips with resistances smaller than the anode surface resistance by a factor of 2. The detector position linearity, position resolution, and energy resolution were measured with alpha-particles and heavy ions. In-beam experimental results reveal a position resolution below 1 mm (FWHM) and a very good non-linearity of less than 1% (rms). The energy resolution determined from 228Th alpha source measurements is around 2% (FWHM).
Explore related subjects
Keep this discovery
A. Banu, Y. Li, M. McCleskey, M. Bullough, S. Walsh, C. A. Gagliardi, L. Trache, R. E. Tribble, C. Wilburn. 2007-12-20. Performance evaluation of novel square-bordered position-sensitive silicon detectors with four-corner readout. https://doi.org/10.1016/j.nima.2008.05.016