Performance of prototypes for the ALICE electromagnetic calorimeter
Abstract
The performance of prototypes for the ALICE electromagnetic sampling calorimeter has been studied in test beam measurements at FNAL and CERN. A $4\times4$ array of final design modules showed an energy resolution of about 11% /$\sqrt{E(\mathrm{GeV})}$ $\oplus$ 1.7 % with a uniformity of the response to electrons of 1% and a good linearity in the energy range from 10 to 100 GeV. The electromagnetic shower position resolution was found to be described by 1.5 mm $\oplus$ 5.3 mm /$\sqrt{E \mathrm{(GeV)}}$. For an electron identification efficiency of 90% a hadron rejection factor of $>600$ was obtained.