Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation

Abstract

The gamma-index test has been commonly adopted to quantify the degree of agreement between a reference dose distribution and an evaluation dose distribution. Monte Carlo (MC) simulation has been widely used for the radiotherapy dose calculation for both clinical and research purposes. The goal of this work is to investigate both theoretically and experimentally the impact of the MC statistical fluctuation on the gamma-index test when the fluctuation exists in the reference, the evaluation, or both dose distributions. To the first order approximation, we theoretically demonstrated in a simplified model that the statistical fluctuation tends to overestimate gamma-index values when existing in the reference dose distribution and underestimate gamma-index values when existing in the evaluation dose distribution given the original gamma-index is relatively large for the statistical fluctuation. Our numerical experiments using clinical photon radiation therapy cases have shown that 1) when performing a gamma-index test between an MC reference dose and a non-MC evaluation dose, the average gamma-index is overestimated and the passing rate decreases with the increase of the noise level in the reference dose; 2) when performing a gamma-index test between a non-MC reference dose and an MC evaluation dose, the average gamma-index is underestimated when they are within the clinically relevant range and the passing rate increases with the increase of the noise level in the evaluation dose; 3) when performing a gamma-index test between an MC reference dose and an MC evaluation dose, the passing rate is overestimated due to the noise in the evaluation dose and underestimated due to the noise in the reference dose. We conclude that the gamma-index test should be used with caution when comparing dose distributions computed with Monte Carlo simulation.

Explore related subjects

Keep this discovery

BibTeXRIS

Yan Jiang Graves, Xun Jia, Steve B. Jiang. 2012-08-11. Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation. https://doi.org/10.1088/0031-9155/58/6/1839

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph