arXiv · 1210.1698
A Rayleigh-Brillouin scattering spectrometer for ultraviolet wavelengths
Abstract
A spectrometer for the measurement of spontaneous Rayleigh-Brillouin scattering line profiles at ultraviolet wavelengths from gas phase molecules has been developed, employing a high-power frequency-stabilized UV laser with narrow bandwidth (2 MHz). The UV light from a frequency-doubled titanium:sapphire laser is further amplified in an enhancement cavity, delivering a 5 Watt UV-beam propagating through the interaction region inside a scattering cell. The design of the RB-scattering cell allows for measurements at gas pressures in the range 0-4 bar and at stably controlled temperatures from -30 to 70 degree Celsius. A scannable Fabry-Perot analyzer with instrument resolution of 232 MHz probes the Rayleigh-Brillouin profiles. Measurements on N2 and SF6 gases demonstrate the high signal-to-noise ratio achievable with the instrument, at the 1% level at the peak amplitude of the scattering profile.
Explore related subjects
Keep this discovery
Ziyu Gu, M. Ofelia Vieitez, Eric-Jan van Duijn, Wim Ubachs. 2012-10-05. A Rayleigh-Brillouin scattering spectrometer for ultraviolet wavelengths. https://doi.org/10.1063/1.4721272
Cite the original work for its findings. Save a collection to share your selection of sources.