Dosimetric and Biologic Differences in Flattened and Flattening-Filter-Free Beam Treatment Plans

Abstract

Purpose: To quantitatively compare the dosimetric and biologic differences in treatment plans from flattened and flattening-filter-free (FFF) beam for three anatomic cancer sites. Methods and Materials: Treatment plans with static intensity-modulated radiotherapy beams and volumetric modulated arc therapy beams were generated for 13 patients for both the flattened beam and the FFF beam of the TrueBeam system. Beam energies of 6 MV and 10 MV were chosen for planning. A total of 104 treatment plans were generated in 13 patients. In order to analyze the biological effectiveness of treatment plans, dose volume histograms (DVH) were utilized. Flattened and FFF beam plans are quantitatively compared. Results: In head and neck cases, for VMAT plans, dose reduction in the FFF beam plans compared to the flattened beam in left cochlea, right submandibular gland and right parotid gland reached up to 2.36 Gy, 1.21 Gy and 1.45 Gy, respectively. Similarly, for static IMRT plans, the dose reduction of the FFF beam plans compared to the flattened beam plans for the same organs reached up to 0.34 Gy, 1.36 Gy and 1.46 Gy, respectively. Overall, for head and neck, the FFF beam plans achieved mean dose reduction of up to 5%, 7% and 9%, respectively for above organs at risk. For lung and prostate cases, the FFF beams provided lower or comparable NTCP values to organ-at-risk (OAR) compared to the flattened beam for all plans. Conclusions: In general, we observed treatment plans utilizing FFF beams can improve dose sparing to OARs without compromising the target coverage. Significant dose sparing effect is obtained for head and neck cancer cases, especially for the cases with relatively large field sizes (about 16x20 cm^2). For lung and prostate cases, compared to the flattened beam, the FFF beam based treatment plans provide lower or comparable dose to most OARs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Yue Yan, Poonam Yadav, Michael Bassetti, Kaifang Du, Daniel Saenz, Paul Harari, Bhudatt R. Paliwal. 2015-06-16. Dosimetric and Biologic Differences in Flattened and Flattening-Filter-Free Beam Treatment Plans. https://arxiv.org/abs/1506.05152

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Voxel-Matching NORDIC: Non-local patch formation by time-series similarity increases tSNR in high-resolution BOLD fMRI

Submillimeter functional magnetic resonance imaging (fMRI) based on blood-oxygenation-level-dependent (BOLD) signal enables the study of brain function at the submillimeter level, uncovering insights into fine-scale organisations like cortical layers and columns. However, its inherently low contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) often limit its reliability and applicability. Noise Reduction with Distribution Corrected Principal Components Analysis (NORDIC PCA) is a locally low-rank denoising algorithm that reduces thermal noise levels in BOLD fMRI in a local patch manner. However, local patches often contain a mixture of signals from multiple tissues that negatively affects the low-rank structure of the patches, which limits the denoising capabilities of the algorithm. We propose an alternative approach for patch formation by gathering similar non-local voxels, dubbed voxel-matching (VM) NORDIC. The results on submillimeter-resolution BOLD fMRI data indicate that VM-NORDIC effectively promotes the low rank of the patches by boosting signal redundancy, allowing for more efficient noise attenuation. Moreover, the method barely affects spatial smoothness due to the non-local voxel selection based on time-series similarity. In particular, VM-NORDIC outperforms standard NORDIC with default local patching (Standard-NORDIC) in terms of temporal SNR (tSNR) (~9-90% larger than Standard-NORDIC; ~23-250% larger than the original) and spatial smoothness estimates (~20% of the smoothness induced by Standard-NORDIC). These improvements are fundamental to improving the validity and precision of fMRI studies at submillimeter resolutions.

physics.med-ph

Computed Tomography Reconstruction Algorithm Using Markov Random Field Model

X-ray computed tomography (CT) reveals the materials' internal structures non-destructively from a tilt series of projected images. Filtered back projection (FBP) is a widely-adopted reconstruction algorithm in CT owing to its small computational cost. Under low-dose or sparse-view conditions, however, FBP often amplifies noise, severely degrading the reconstructed images. In this study, we evaluated the performance of a Bayesian CT reconstruction algorithm based on the Markov random field model under such adverse conditions. Through simulations, we demonstrated that the proposed algorithm shows higher reconstruction performance than FBP under both low-dose and sparse-view conditions. The hyperparameters are estimated by minimizing the Bayesian free energy, enabling adaptive reconstruction that reflects the noise characteristics of the observed projection data. These results suggest that the proposed algorithm can broaden the applicability of CT to dose-sensitive applications and time-constrained measurements, where only limited observed projection data are available.

physics.med-ph

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion--ion recombination and self-consistent space-charge electric field calculations. The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions. The implementation was validated against analytical and numerical models from the literature. Excellent agreement was obtained for the free electron fraction (FEF), CCE, induced current, and electric field evolution under UHDR conditions. The simulations show that space charge can locally increase the electric field by more than a factor of four or reduce it to nearly zero in some regions of the chamber. The results further suggest that the reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment. This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.

physics.med-ph