Abstract
The ability of background discrimination using pulse shape discrimination (PSD) in broad-energy germanium (BEGe) detectors makes them as competitive candidates for neutrinoless double beta decay (0{\nu}\b{eta}\b{eta}) experiments. The measurements of key parameters for detector modeling in a commercial p-type BEGe detector are presented in this paper. Point-like sources were used to investigate the energy resolution and linearity of the detector. A cylindrical volume source was used for the efficiency calibration. With an assembled device for source positioning, a collimated 133Ba point-like source was used to scan the detector and investigate the active volume. A point-like source of 241Am was used to measure the dead layer thicknesses, which are approximately 0.17 mm on the front and 1.18 mm on the side. The described characterization method will play an important role in the 0{\nu}\b{eta}\b{eta} experiments with BEGe detectors at China JinPing underground Laboratory (CJPL) in the future.
Explore related subjects
Keep this discovery
Zhi Zeng, Yu-Hao Mi, Ming Zeng, Hao Ma, Qian Yue, Jian-Ping Cheng, Jun-Li Li, Rui Qiu, Hui Zhang. 2016-03-06. Characterization study of a broad-energy germanium detector at CJPL. https://arxiv.org/abs/1603.01782
Cite the original work for its findings. Save a collection to share your selection of sources.