Search arXivSearch

arXiv · 1703.02448

Non-contact luminescence lifetime cryothermometry for macromolecular crystallography

Abstract

A novel technique for remote, non-contact, in situ monitoring of the protein crystal temperature has been developed for the new I23 beamline at the Diamond Light Source, a facility dedicated to macromolecular crystallography (MX) with long-wavelength X-rays. The temperature is derived from the temperature-dependant decay time constant of luminescence from a minuscule scintillation sensor (0.05 mm3) located in very close proximity to the sample under test. In this work we present the underlying principle of cryogenic luminescence lifetime thermometry, discuss the features of the detection method, the choice of temperature sensor and demonstrate how the temperature monitoring system was integrated within the viewing system of the end-station used for the visualisation of protein crystals. The thermometry system was characterised using a Bi4Ge3O12 (BGO) crystal scintillator that exhibits good responsivity of the decay time constant as function of temperature over a wide range (8-270 K). The scintillation sensor was calibrated and the uncertainty of the temperature measurements over the primary operation temperature range of the beamline (30-150 K) was assessed to be +-1.6 K. It has been shown that the temperature of the sample holder, measured using the luminescence sensor, agrees well with the expected value. The technique was applied to characterise the thermal performance of different sample mounts that have been used in MX experiments at the I23 beamline. The thickness of the mount is shown to have the greatest impact upon the temperature distribution across the sample mount. Altogether these tests and findings demonstrate the usefulness of the thermometry system in highlighting the challenges that remain to be addressed for the in-vacuum MX experiment to become a reliable and indispensable tool for structural biology

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

V. Mykhaylyk A. Wagner, H. Kraus. 2017-03-07. Non-contact luminescence lifetime cryothermometry for macromolecular crystallography. https://doi.org/10.1107/s1600577517003484

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det