arXiv · 1707.03392
On the use of NaI scintillation for high stability nuclear decay rate measurements
Abstract
We demonstrate the linearity and stability of a gamma-ray scintillation detector comprised of a NaI(Tl) crystal and a scientific-grade CMOS camera. After calibration, this detector exhibits excellent linearity more than three decades of activity levels. Because the detector is not counting pulses, no dead-time correction is required. When high activity sources are brought into close proximity to the NaI crystal, several minutes are required for the scintillation to achieve a steady state. On longer time scales, we measure drifts of a few percent over several days. These instabilities have important implications for precision determinations of nuclear decay rate stability.
Explore related subjects
Keep this discovery
Scott D. Bergeson, Michael J. Ware, Jeremy Hawk. 2017-07-10. On the use of NaI scintillation for high stability nuclear decay rate measurements. https://arxiv.org/abs/1707.03392