Search arXivSearch

arXiv · 1710.10972

Development of Thermal Kinetic Inductance Detectors suitable for X-ray spectroscopy

Abstract

We report on the development of Thermal Kinetic Inductance Detectors (TKIDs) suitable to perform X-ray spectroscopy measurements. The aim is to implement MKIDs sensors working in thermal quasi-equilibrium mode to detect X-ray photons as pure calorimeters. The thermal mode is a variation on the MKID classical way of operation that has generated interest in recent years. TKIDs can offer the MKIDs inherent multiplexibility in the frequency domain, a high spatial resolution comparable with CCDs, and an energy resolution theoretically limited only by thermodynamic fluctuations across the thermal weak links. Microresonators are built in Ti/TiN multilayer technology with the inductive part thermally coupled with a metal absorber on a suspended SiN membrane, to avoid escape of phonons from the film to the substrate. The mid-term goal is to optimize the single pixel design in term of superconducting critical temperatures, internal quality factors, kinetic inductance and spectral energy resolution. The final goal is to realize a demonstrator array for a next generation thousand pixels X-ray spectrometer. In this contribution, the status of the project after one year of developments is reported, with detailed reference to the microresonators design and simulations and to the fabrication process.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

A. Giachero, A. Cruciani, A. D'Addabbo, P. K. Day, S. Di Domizio, M. Faverzani, E. Ferri, B. Margesin, M. Martinez, R. Mezzena, L. Minutolo, A. Nucciotti, A. Puiu, M. Vignati. 2017-10-30. Development of Thermal Kinetic Inductance Detectors suitable for X-ray spectroscopy. https://doi.org/10.1007/s10909-018-2043-8

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det