A novel ppm-precise absolute calibration method for precision high-voltage dividers

Abstract

The most common method to measure direct current high voltage (HV) down to the ppm-level is to use resistive high-voltage dividers. Such devices scale the HV into a range where it can be compared with precision digital voltmeters to reference voltages sources, which can be traced back to Josephson voltage standards. So far the calibration of the scale factors of HV dividers for voltages above 1~kV could only be done at metrology institutes and sometimes involves round-robin tests among several institutions to get reliable results. Here we present a novel absolute calibration method based on the measurement of a differential scale factor, which can be performed with commercial equipment and outside metrology institutes. We demonstrate that reproducible measurements up to 35~kV can be performed with relative uncertainties below $1\cdot10^{-6}$. This method is not restricted to metrology institutes and offers the possibility to determine the linearity of high-voltage dividers for a wide range of applications.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

O. Rest, D. Winzen, S. Bauer, R. Berendes, J. Meisner, T. Thümmler, S. Wüstling, C. Weinheimer. 2019-02-28. A novel ppm-precise absolute calibration method for precision high-voltage dividers. https://doi.org/10.1088/1681-7575%2Fab2997

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det