ARRAY: An Open Source, Modular and Probe-Card based System with Integrated Switching Matrix for Characterisation of Large Area Silicon Pad Sensors

Abstract

Silicon pad sensors are proposed as active material in highly granular sampling calorimeters of future collider experiments such as the Compact Linear Collider (CLIC) or the International Linear Collider (ILC). The electromagnetic section of these designs often include O(1000 m$^2$) of silicon pad sensors. For the luminosity measurement, a dedicated forward calorimeter called LumiCal is foreseen. More recently, the CMS experiment has decided to adopt the same concept in its endcap calorimeter upgrade for the HL-LHC. The sensors are typically produced from 6- or 8-inch wafers and consist of a few hundred smaller cells, each with an area of O(0.1 to 1 $\text{cm}^2$). For the prototyping phase of these projects, several design choices have to be evaluated while for mass production, thousands of sensors have to be tested for quality control. For the electrical characterisation of these sensors, it is important to bias them under realistic conditions. To fulfil these requirements, ARRAY, a compact, modular and cost efficient system for large area silicon pad sensor characterisation has been developed and successfully commissioned. It consists of two plugin printed circuit boards: an active switching matrix with 512 input channels that holds all controls and a passive probe card that connects to the sensor. The latter can then be adapted to any sensor geometry. All design files are open source. The system has been used to measure currents ranging from 500 pA to 5 $μ$A and capacitances between 5 pF and 100 pF. A precision of better than 0.2 pF on capacitance measurements in that range can be achieved. Examples of calibration and measurement results for leakage current and capacitance are presented.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Erica Brondolin, Dominik Dannheim, Szymon Kulis, Andreas A. Maier, Florian Pitters, Thorben Quast, Eva Sicking. 2021-02-26. ARRAY: An Open Source, Modular and Probe-Card based System with Integrated Switching Matrix for Characterisation of Large Area Silicon Pad Sensors. https://doi.org/10.1016/j.nima.2019.06.007

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

TCAD + Allpi$\text{x}^2$ Simulation study of MALTA2, a Depleted Monolithic Active Pixel Sensor for future tracking

In this work, a hybrid simulation framework combining TCAD and Allpi$\text{x}^2$ is presented to investigate the sensor properties of MALTA2, a depleted monolithic active pixel sensor designed for future tracking. The study starts from 3D modeling and transient simulations in TCAD, with generic doping profiles and simple well structures. The resulting doping profiles and electric field are extracted and fed into Allpi$\text{x}^2$ for high-statistics Monte Carlo simulations in both DUT-only and full-telescope mode. Simulations reveal a strong dependence of sensor performance, specifically the detection efficiency and cluster size, on the doping concentration of the N-type blanket at the sensor surface. The doping concentration is then optimized by comparing simulations with measurement data. The active depth of the depleted region of the MALTA2 sensor is estimated in both simulations and measurements using a grazing angle method, in which the sensor is positioned at various inclinations relative to the beam, covering angles from 0 to 60 degrees. Excellent agreement on active depth is obtained with the optimal doping concentration, showing a deviation of 2\% from the measured value at a threshold of 450\,$\text{e}^-$. Consequently, the framework offers a generic toolkit for sensor studies without requiring proprietary information.

physics.ins-det

A compact fibre-coupled active beamstop for fast time-resolved monitoring of synchrotron X-ray beams

We have developed a compact fibre-coupled active beamstop for simultaneous interception and time-resolved monitoring of the direct X-ray beam in synchrotron diffraction experiments. The complete beamstop tip has a maximum transverse dimension of 1 mm and incorporates a 700 $μ$m-wide Ce:YAG scintillator. The scintillator converts the intercepted X-rays into visible light, which is transported through a multimode optical fibre to a silicon photodiode at the bottom of the blade that is connected to a dedicated fast amplifier electronics. Tests with a pulsed laser yielded a system decay time of 69.4 ns, consistent with reported luminescence decay times for Ce:YAG. Measurements at the P14.EH2 end station of PETRA III demonstrated the detection of individual bunch signals separated by 192 ns in the 40-bunch operating mode. The active beamstop also provided an in-situ timing signal for synchronising an experimental sequence with the approximately 190 $μ$s transmission window of a mechanical X-ray chopper. The device therefore combines a compact obstruction of the direct beam with fast monitoring of X-ray timing and flux during diffraction data acquisition.

physics.ins-det

Charge-Carrier transport simulations in diamond detectors with electric-field-dependent mobility and charge-collection-distance-based trapping

Diamond detectors are attractive for operation in harsh radiation environments because they combine radiation tolerance, fast signal formation, and low leakage current. Realistic detector-response simulations require an accurate description of charge-carrier mobility and trapping, which determine both signal amplitude and timing. In this work, we extend \allpix{}, a modular end-to-end detector simulation framework, with diamond-specific transport models. The implementation includes field-dependent mobility parameterizations for electrons and holes and an effective trapping model based on the charge collection distance (CCD), providing a detector-level interface to material quality and radiation-damage measurements. The mobility description is validated in the negligible-trapping limit using single-crystalline CVD diamond by comparing simulated drift velocities and transient-current signals with published reference data. For polycrystalline CVD diamond, the CCD-based trapping model is evaluated using experimentally measured CCD values and compared with laboratory transient-current-technique waveforms. The simulations reproduce the measured drift-velocity behavior in scCVD and the reduced charge collection and degraded transient response observed in pcCVD. The presented implementation enables detector-level studies of charge collection, pulse formation, and timing performance in diamond sensors using experimentally accessible transport and trapping parameters, and provides a practical framework for simulation-driven detector development and radiation-damage studies.