arXiv · 1911.01883
Compensation of phase drifts caused by ambient humidity, temperature and pressure changes for continuously operating interferometers
Abstract
Fusion experiments rely heavily on the measurement of the line-integrated electron density by interferometry for density feed-back control. In recent years the discharge length has increased dramatically and is continuing to rise, resulting in environmentally induced phase drifts to become an increasingly worrisome subject, since they falsify the interferometer's measurement of the density. Especially in larger Tokamaks the loss of density control due to uncontrolled changes in the optical path length can have a disastrous outcome. The control of environmental parameters in large diagnostic/experimental halls is costly and sometimes infeasible and in some cases cannot be retro-fitted to an existing machine. In this report we present a very cheap (ca. 100 EUR), easily retro-fitted, real-time capable phase compensation scheme for interferometers measuring dispersive media over long time scales. The method is not limited to fusion, but can be applied to any continuously measuring interferometer measuring a dispersive medium. It has been successfully applied to the Wendelstein 7-X density feed-back interferometer.
Explore related subjects
Keep this discovery
K. J. Brunner, J. Knauer, J. Meineke, M. Stern, M. Hirsch, B. Kursinski, R. C. Wolf, the W7-X team. 2019-11-05. Compensation of phase drifts caused by ambient humidity, temperature and pressure changes for continuously operating interferometers. https://doi.org/10.1088/1748-0221%2F14%2F11%2Fp11016
Cite the original work for its findings. Save a collection to share your selection of sources.