Characterization of High Temperature Calibration Bath through Stability and Uniformity Tests with Data Acquisition using Standard Platinum Resistance Thermometer and Precision Multimeter

Abstract

Calibration is one area in measurement where our country is still in the process of improvement. The absence of confirmed approaches and actions for identifying the features of calibration baths by numerous laboratories disallowed them to assess more the finest measurement ability and develop their accurateness. In recent advances in technology, there is now a way of providing the best measurement capability in temperature calibration by providing validated methods in characterizing calibration baths that contributes to the uncertainty of measurement of the calibrated thermometer and is therefore important to measure the extent of its contribution to the final uncertainty. This study adapted the process from obtainable and current methods. Platinum Resistance Thermometer (PRT-Isotech) and Standard Platinum Resistance Thermometer (SPRT-Hart) which are interfaced to Precision Digital Multimeters were used for measurements. Quite a few measurements were made at varying time durations implemented on several known positions in an orderly measurement form which enclosed the whole workspace. The uniformities and stabilities of the calibration bath were computed by getting the range of the Minimum Difference with the Maximum Difference taken at every established temperature point. Outcomes of measurements exhibited that the calibration bath stability at every temperature setting are steadier at lesser temperatures and have a tendency to rise as the temperature approaches 250oC. The liquid bath was further unvarying at lesser temperatures. The variances found were insignificantly distributed since the Standard Deviation ranged only from 0.005 to 0.013.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Arlene G. Estacio, Vilma C. Pagtalunan, Rio S. Pagtalunan, Ira C. Valenzuela, Lean Karlo S. Tolentino, Jennifer C. Dela Cruz. 2019-11-13. Characterization of High Temperature Calibration Bath through Stability and Uniformity Tests with Data Acquisition using Standard Platinum Resistance Thermometer and Precision Multimeter. https://doi.org/10.30534/ijeter%2F2019%2F047102019

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det