arXiv · 2008.11860
New Method for Silicon Sensor Charge Calibration Using Compton Scattering
Abstract
In order to cope with increasing lifetime radiation damage expected at collider experiments, silicon sensors are becoming increasingly thin. To achieve adequate detection efficiency, the next generation of detectors may have to operate with thresholds below 1000 electron-hole pairs. The readout chips attached to these sensors should be calibrated to some known external charge, but there is a lack of traditional sources in this charge regime. We present a new method for absolute charge calibration based on Compton scattering. In the past, this method has been used for calibration of scintillators, but to our knowledge never for silicon detectors. Here it has been studied using a 150 micron thick planar silicon sensor on an RD53A readout integrated circuit.
Explore related subjects
Keep this discovery
Patrick McCormack, Maurice Garcia-Sciveres, Timon Heim, Benjamin Nachman, Magne Lauritzen. 2020-08-26. New Method for Silicon Sensor Charge Calibration Using Compton Scattering. https://arxiv.org/abs/2008.11860
Cite the original work for its findings. Save a collection to share your selection of sources.