arXiv · 2102.05942
Improvements on INRIM Coaxial Microcalorimeter and Outcome of a Model Comparison
Abstract
This paper describes hardware and software improvements of the INRIM coaxial microcalorimeter together with their outcome on the primary power standard realization in the frequency band 0.05 - 40 GHz. A better temperature and power stabilization turned out to provide an improved signal/noise ratio and a drift reduction in every working condition of the microcalorimeter. The INRIM correction model is also compared to a traditional, but faster, one in terms of measurement uncertainty. Outcomes are presented in form of a 2.92 mm thermoelectric power sensor calibration together with results that show the improved stability and repeatability of the measurement system.
Explore related subjects
Keep this discovery
Luciano Brunetti, Luca Oberto, Marco Sellone, Nosherwan Shoaib, Emil Vremera. 2021-02-11. Improvements on INRIM Coaxial Microcalorimeter and Outcome of a Model Comparison. https://doi.org/10.1109/tim.2015.2406052