arXiv · 2105.02178
Diagnostics for ultrashort X-ray pulses using silicon trackers
Abstract
The spectrum of laser-plasma generated X-rays is very important, it characterizes electron dynamics in plasma and is basic for applications. However, the accuracies and efficiencies of existing methods to diagnose the spectrum of laser-plasma based X-ray pulse are not very high, especially in the range of several hundred keV. In this study, a new method based on electron tracks detection to measure the spectrum of laser-plasma produced X-ray pulses is proposed and demonstrated. Laser-plasma generated X-rays are scattered in a multi-pixel silicon tracker. Energies and scattering directions of Compton electrons can be extracted from the response of the detector, and then the spectrum of X-rays can be reconstructed. Simulations indicate that the energy resolution of this method is approximately 20% for X-rays from 200 to 550 keV for a silicon-on-insulator pixel detector with 12 $\rm \mu$m pixel pitch and 500 $\rm \mu$m depletion region thickness. The results of a proof-of-principle experiment based on a Timepix3 detector are also shown.
Explore related subjects
Keep this discovery
Jiaxing Wen, Minghai Yu, Yuchi Wu, Ming Zeng, Bo Zhang, Jirong Cang, Yuge Zhang, Ge Ma, Yue Yang, Wenbo Mo, Zongqing Zhao. 2021-05-03. Diagnostics for ultrashort X-ray pulses using silicon trackers. https://doi.org/10.1016/j.nima.2021.165754
Cite the original work for its findings. Save a collection to share your selection of sources.