Optimization of FLASH Proton Beams Using a Track-Repeating Algorithm

Abstract

Methods: A phase space file in a plane at 202 mm downstream of the beam exit window is generated through tuning parameters to match FDC results with measured or MCNPX Monte Carlo-simulated integrated depth-dose distribution (IDD) and lateral dose profiles. To spread out the Bragg peak, widen the beam and reduce the penumbra, a ridge filter (RF), a high-Z material scatterer and a collimator with compensator are inserted in the beam path and their shapes and sizes have been optimized. The FDC calculations are validated by comparing Geant4 Monte Carlo simulations. In addition, a set of algorithms to automatically choose the optimum dimensions of the beam shaping elements is developed and tested using the same beams. At the last part, dose rates for optimized beams were estimated by scaling their dose distributions to that of their original beams. Results: The optimized 86.4 MeV beam had an 8.5 mm wide spread-out Bragg peak (SOBP) (proximal 90% to distal 90% of the maximum dose), 14.5 mm, 12.0 mm and 11.0 lateral widths with dose above 50%, 80% and 90% respectively and a 2.5 mm penumbra from 80% to 20% in the lateral profile for the energy. The 159.5 MeV beam had a SOBP of 39.0 mm and the lateral widths with dose above 50%, 80% and 90% of 20.5 mm, 15.0 and 12.5 mm when the source to surface distance (SSD) was 550 mm. Wider lateral widths was obtained with increased SSD. The FDC calculations had passing rates higher than 96% using 3mm/3% as the gamma-index criterion comparing with Geant4 simulations for both energies. The set of automatic algorithms can choose the proper dimensions for the high-density scatterer, RF, collimator and compensator efficiently. And the optimized 159.5 MeV beam with different SDDs had entrance dose rate higher than 40 Gy/s if the entrance dose rate of the original beam was 150 Gy/s.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Qianxia Wang, Uwe Titt, Radhe Mohan, Fada Guan, Yao Zhao, Ming Yang, Pablo Yepes. 2021-11-30. Optimization of FLASH Proton Beams Using a Track-Repeating Algorithm. https://doi.org/10.1002/mp.15849

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Voxel-Matching NORDIC: Non-local patch formation by time-series similarity increases tSNR in high-resolution BOLD fMRI

Submillimeter functional magnetic resonance imaging (fMRI) based on blood-oxygenation-level-dependent (BOLD) signal enables the study of brain function at the submillimeter level, uncovering insights into fine-scale organisations like cortical layers and columns. However, its inherently low contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) often limit its reliability and applicability. Noise Reduction with Distribution Corrected Principal Components Analysis (NORDIC PCA) is a locally low-rank denoising algorithm that reduces thermal noise levels in BOLD fMRI in a local patch manner. However, local patches often contain a mixture of signals from multiple tissues that negatively affects the low-rank structure of the patches, which limits the denoising capabilities of the algorithm. We propose an alternative approach for patch formation by gathering similar non-local voxels, dubbed voxel-matching (VM) NORDIC. The results on submillimeter-resolution BOLD fMRI data indicate that VM-NORDIC effectively promotes the low rank of the patches by boosting signal redundancy, allowing for more efficient noise attenuation. Moreover, the method barely affects spatial smoothness due to the non-local voxel selection based on time-series similarity. In particular, VM-NORDIC outperforms standard NORDIC with default local patching (Standard-NORDIC) in terms of temporal SNR (tSNR) (~9-90% larger than Standard-NORDIC; ~23-250% larger than the original) and spatial smoothness estimates (~20% of the smoothness induced by Standard-NORDIC). These improvements are fundamental to improving the validity and precision of fMRI studies at submillimeter resolutions.

physics.med-ph

Computed Tomography Reconstruction Algorithm Using Markov Random Field Model

X-ray computed tomography (CT) reveals the materials' internal structures non-destructively from a tilt series of projected images. Filtered back projection (FBP) is a widely-adopted reconstruction algorithm in CT owing to its small computational cost. Under low-dose or sparse-view conditions, however, FBP often amplifies noise, severely degrading the reconstructed images. In this study, we evaluated the performance of a Bayesian CT reconstruction algorithm based on the Markov random field model under such adverse conditions. Through simulations, we demonstrated that the proposed algorithm shows higher reconstruction performance than FBP under both low-dose and sparse-view conditions. The hyperparameters are estimated by minimizing the Bayesian free energy, enabling adaptive reconstruction that reflects the noise characteristics of the observed projection data. These results suggest that the proposed algorithm can broaden the applicability of CT to dose-sensitive applications and time-constrained measurements, where only limited observed projection data are available.

physics.med-ph

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion--ion recombination and self-consistent space-charge electric field calculations. The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions. The implementation was validated against analytical and numerical models from the literature. Excellent agreement was obtained for the free electron fraction (FEF), CCE, induced current, and electric field evolution under UHDR conditions. The simulations show that space charge can locally increase the electric field by more than a factor of four or reduce it to nearly zero in some regions of the chamber. The results further suggest that the reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment. This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.

physics.med-ph