EMIR, the near-infrared camera and multi-object spectrograph for the GTC

Abstract

We present EMIR, a powerful near-infrared (NIR) camera and multi-object spectrograph (MOS) installed at the Nasmyth focus of the 10.4 m GTC. EMIR was commissioned in mid-2016 and is offered as a common-user instrument. It provides spectral coverage of 0.9 to 2.5 $μm$ over a field of view (FOV) of 6.67x6.67 squared arcmin in imaging mode, and 6.67x4 squared arcmin in spectroscopy. EMIR delivers up to 53 spectra of different objects thanks to a robotic configurable cold slit mask system that is located inside the cryogenic chamber, allowing rapid reconfiguration of the observing mask. The imaging mode is attained by moving all bars outside the FOV and then leaving an empty space in the GTC focal surface. The dispersing suite holds three large pseudo-grisms, formed by the combination of high-efficiency FuSi ion-etched ruled transmission grating sandwiched between two identical ZnSe prisms, plus one standard replicated grism. These dispersing units offer the spectral recording of an atmospheric window $J,H,K$ in a single shot with resolving powers of 5000, 4250, 4000, respectively for a nominal slit width of 0.6\arcsec, plus the combined bands $YJ$ or $HK$, also in a single shot, with resolution of $\sim$ 1000. The original Hawaii2 FPA detector, which is prone to instabilities that add noise to the signal, is being replaced by a new Hawaii2RG detector array, and is currently being tested at the IAC. This paper presents the most salient features of the instrument, with emphasis on its observing capabilities and the functionality of the configurable slit unit. Sample early science data is also shown.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F. Garzón, M. Balcells, J. Gallego, C. Gry, R. Guzmán, P. Hammersley, A. Herrero, C. Muñoz-Tuñón, R. Pelló, M. Prieto, É. Bourrec, C. Cabello, N. Cardiel, C. González-Fernández, N. Laporte, B. Milliard, S. Pascual, L. R. Patrick, J. Patrón, S. Ramírez-Alegría, A. Streblyanska. 2022-09-30. EMIR, the near-infrared camera and multi-object spectrograph for the GTC. https://doi.org/10.1051/0004-6361%2F202244729

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Performance characterization of a new Structural and Thermal Architecture for a future spaceborne Closed-Cycle Dilution Refrigerator

A Structural and Thermal Model (STM) has been developed to support the new spaceborne Closed-Cycle Dilution Refrigerator (CCDR), which aims to provide continuous cooling at 100~mK for long-duration astrophysical missions. The STM is based on a hexapod architecture that ensures both thermal decoupling and mechanical robustness during launch. In this paper, we present the characterization of its thermal and mechanical performances. A dedicated experimental setup was used to investigate the thermal behavior of the STM across a broad temperature range. The study reveals limitations of the collar design, with incomplete power interception from thermal boundary resistances and vibration test failure traced to defective strut gluing. These results guide the next STM iteration with optimized collar and strut assembly for reliable CCDR operation in space.

astro-ph.IM

The Simons Observatory: Development of a Pipeline to Detect Rapid Transients in Time-Ordered Data

We introduce a method for detecting astrophysical transients evolving on timescales of milliseconds to minutes using cosmic microwave background (CMB) survey telescopes. While previous transient searches in CMB data operate in map space, our pipeline directly processes the raw time-ordered data, enabling sensitivity to fast, dynamic signals. We integrate our detection approach into the Simons Observatory time-domain pipeline and assess the performance by injecting symmetric, stellar flare-like light curves into simulated observations. For events flaring with a timescale of 0.5 s, the pipeline detects $\gtrsim90$ % of events at flux densities of 800, 1150, 1650, and 4250\,mJy when measured in the 93, 145, 225, and 280 GHz bands respectively. At a fixed peak flux density, the pipeline more readily detects longer flares. The limiting flux density for 90 % completeness is four times lower for a $\ge5$ s flare than for a 0.5 s flare, while the flux density limits for $\gtrsim50$ % detection efficiency are comparable to the rms noise of the time-ordered data. We are able to determine the position of detected events in each observing band, with a positional uncertainty at the detection threshold comparable to the telescope resolution at that band. These results demonstrate the readiness of this pipeline for incorporation into upcoming Simons Observatory data analyses.

astro-ph.IM

Thermal conductivity of various CFRPs from 100 mK to 20 K

Carbon-fiber-reinforced polymers (CFRPs) are some of the most useful materials for building spacecraft and aerospace tools. They are especially valuable for systems that work at extremely cold (cryogenic) temperatures because they are strong, lightweight, and don't transfer heat easily. In this study, researchers measured how well heat moves through several different types of carbon fiber samples, specifically T300, T700, HS40, M55J, and IMA, at different fiber layouts and densities. These measurements were taken at ultra-cold temperatures ranging from 100 mK to 20 K. The team used a newly developed analysis method to calculate the thermal conductivity for each sample. Finally, they shared how each material behaved at different temperatures and compared their findings to previous research.

astro-ph.IM