Search arXivSearch

arXiv · 2308.02545

Open-access microcavities: high stability without dedicated mechanical low-pass filter in closed-cycle cryostats

Abstract

Open-access optical microcavities are Fabry-Perot type cavities consisting of two micrometer-size mirrors, separated by an air (or vacuum) gap typically of a few micrometers. Compared to integrated microcavities, this configuration is more flexible as the relative position of the two mirrors can be tuned, allowing to change on demand parameters such as cavity length and mode volume, and to select specific transverse cavity modes. These advantages come at the expense of mechanical stability of the cavity itself, which is particularly relevant in noisy closed-cycle cryostats. Here we show an open-access optical microcavity based on scanning-probe microscope design principles. When operated at 4 K in a tabletop optical closed-cycle cryostat without any dedicated low-pass filter, we obtain stabilities of 5.7 and 10.6 pm rms in the quiet and full period of the cryocooler cycle, respectively. Our device has free-space optical access, essential for instance for full polarization control.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M. Fisicaro, M. J. Rost, M. Witlox, H. van der Meer, W. Löffler. 2023-08-01. Open-access microcavities: high stability without dedicated mechanical low-pass filter in closed-cycle cryostats. https://arxiv.org/abs/2308.02545

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det