Particle Tracking, Recognition and LET Evaluation of Out-of-Field Proton Therapy Delivered to a Phantom with Implants

Abstract

This study aims to assess the composition of scattered particles generated in proton therapy for tumours situated proximal to titanium dental implants. The investigation involves decomposing the mixed field and recording Linear Energy Transfer (LET) spectra to quantify the influence of metallic dental inserts located behind the tumour. A conformal proton beam was used to deliver the treatment plan to an anthropomorphic head phantom with two types of implants (Ti and plastic) inserted in the target volume. The stray radiation resulting during the irradiation was detected by a hybrid semiconductor pixel detector MiniPIX Timepix3 that was placed distal to the Spread-out Bragg peak. Visualization and field decomposition of stray radiation were generated using algorithms trained in particle recognition based on artificial intelligence convolution neural networks (AI CNN). Spectral sensitive aspects of the scattered radiation were collected using two angular positions of the detector relative to the beam direction: 0 and 60°. Using AI CNN, 3 classes of particles were identified: protons, electrons & photons, ions & fast neutrons. Placing a Ti implant in the beam's path resulted in predominantly electrons and photons, contributing 52.2%, whereas for plastic implants, the contribution was 65.4%. Scattered protons comprised 45.5% and 31.9% with and without Ti inserts, respectively. The LET spectra was derived for each group of particles, with values ranging from 0.01 to 7.5 keVμm-1 for Ti/plastic implants. The low-LET component was primarily composed of electrons and photons, while the high-LET component corresponded to protons and ions. This method, complemented by directional maps, holds potential for evaluating and validating treatment plans involving stray radiation near organs at risk, offering precise discrimination of the mixt field, enhancing in this way the LET calculation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Cristina Balan, Carlos Granja, Gennady Mytsin, Sergey Shvidky, Alexander Molokanov, Lukas Marek, Vasile Chis, Cristina Oancea. 2023-12-21. Particle Tracking, Recognition and LET Evaluation of Out-of-Field Proton Therapy Delivered to a Phantom with Implants. https://arxiv.org/abs/2312.14304

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Explicit pixel-value correction of lateral response artifacts in radiochromic film dosimetry: Evaluation of single-, double-, and triple-channel methods

Background:Flatbed scanners introduce a lateral response artifact (LRA) perpendicular to lamp motion. Because LRA varies with position, pixel value (PV), color channel, and film model, it may violate the channel-common perturbation assumption of multichannel dosimetry. Purpose:To characterize PV-domain LRA in EBT4 and EBT-XD films, compare Lewis and Full corrections, and evaluate their effects on single-, double-, and triple-channel dosimetry. Methods:Seven films were scanned at seven lateral positions using a cyclic-shift design. For each position and RGB channel, center-equivalent response was modeled as PVc = A + B * PV. Lewis correction used two films with linear interpolation of A and B; Full correction used all center-local pairs with PCHIP interpolation of PVc - PV. Performance was assessed by portrait-landscape profile consistency and dose accuracy. Results: LRA was affine in PV, asymmetric about scanner center, and dependent on film model and channel. Both corrections reduced profile differences. Mean absolute profile differences across MU levels ranged from 1.4%-5.3% for EBT4 and 2.4%-6.8% for EBT-XD. The methods performed similarly for EBT4, while Full correction reduced discrepancies for several EBT-XD double- and triple-channel methods. Residual blue-channel errors limited blue, red-blue, and triple-channel performance. At 500 MU and above, red showed the best dose agreement; at 100 MU, both corrections increased deviations for several single-channel estimates. Full correction produced smaller deviations than Lewis in several EBT-XD conditions. Conclusions:LRA should be corrected in the raw-PV domain before dose reconstruction. Full correction better preserved profile consistency and dose accuracy, particularly for EBT-XD. Because improved profile consistency did not always improve central-dose agreement, both should be evaluated separately in commissioning.

physics.med-ph

Integration of Spectral CT with PET and SPECT: Bringing Tissue Composition Information to Molecular Imaging

Molecular imaging has been transformed by integrating positron emission tomography (PET) or single-photon emission computed tomography (SPECT) with x-ray computed tomography (CT). However, x-ray CT in hybrid imaging is used primarily for anatomical localization and for attenuation and scatter correction of emission data. Its ability to characterize tissue composition remains underutilized. Spectral CT, including dual-energy CT (DECT) and photon-counting CT (PCCT), can provide material-specific information, such as images of iodine concentration, bone or calcium fractions, electron-density, effective atomic number and other spatially-varying material properties. These capabilities create an opportunity to bring tissue-composition information into molecular imaging. In addition, spectral CT can also reduce CT artifacts and provide more accurate corrections for emission data. In this review, we discuss the technical basis, integration pathways, and translational opportunities for combining spectral CT with molecular imaging. We review sequential and integrated PET/DECT approaches, emerging PET/PCCT concepts, PET-enabled spectral CT, and extensions to SPECT/CT. We then examine how spectral CT information may improve attenuation correction, scatter correction, tissue-fraction correction, positron range correction, reconstruction priors, and dosimetry. Clinical and translational applications include contrast-enhanced molecular imaging, vascular and perfusion imaging, quantitative bone marrow imaging, musculoskeletal imaging, and theranostics. Overall, spectral CT integration may move hybrid molecular imaging toward tissue-composition-informed molecular imaging, in which radiotracer signals are interpreted within their material context.

physics.med-ph

Non-circular scan trajectories for reducing cone-beam artifacts in Gamma Knife CBCT images: a simulation study

Objective. Gamma Knife cone-beam computed tomography (CBCT) images are deteriorated by cone-beam artifacts whose magnitude increases along the superior direction. In this study, a novel scan trajectory compatible with the Gamma Knife CBCT system is optimized to reduce cone-beam artifacts, with the aim to replace the current 200-degree single-arc scan. Approach. Data sampling analysis with tomographic incompleteness maps indicates the level of undersampling across the field of view based on the geometry of the system and of a scan trajectory. Moreover, CBCT simulations are performed from a virtual phantom with disks aligned along the axial direction and from a CT reconstruction of a stereotactic end-to-end validation (STEEV) phantom. CBCT projections are simulated for a given scan trajectory through a polychromatic forward projection model with added noise and scatter, then the CBCT image is reconstructed using an iterative algorithm which minimizes weighted least squares. Main results. Both the incompleteness analysis and the CBCT simulations indicate adding lines to the current single-arc trajectory is more efficient to reduce cone-beam artifacts than adding more arcs, both in terms of number of additional projections and new artifacts. A line-arc-line trajectory with source axial steps of 3.5 mm removes virtually all cone-beam artifacts. The widths of the cone-beam artifacts created by the disks show a positive correlation between the artifact magnitude and the incompleteness value. Significance. A line-arc-line scan trajectory is promising to reduce cone-beam artifacts of the Gamma Knife CBCT images while being a compatible and reasonable trajectory for the current system design.

physics.med-ph