arXiv · 2402.03344
First (calibration) experiment using proton beam from FRENA at SINP
Abstract
This work presents the first calibration experiment of a 3 MV Tandetron accelerator, FRENA, performed in May 2022. The $^7$Li(p,n) reaction threshold was measured to calibrate the terminal voltage measuring device. A LiF target of thickness 175 $\mu$g/cm$^2$ was used in the experiment. The measured threshold was 1872$\pm$2.7 keV, indicating 6$-$10 keV energy shift.
Explore related subjects
Keep this discovery
C. Basu, K. Banerjee, T. K. Ghosh, G. Mukherjee, C. Bhattacharya, Shraddha S Desai, R. Shil, A. K. Saha, J. K. Meena, T. Bar, D. Basak, L. K. Sahoo, S. Saha, C. Marick, D. Das, M. Kujur, S. Roy, S. S. Basu, U. Gond, A. Saha, A. Das, M. Samanta, P. Saha, S. K. Karan. 2024-01-24. First (calibration) experiment using proton beam from FRENA at SINP. https://arxiv.org/abs/2402.03344