A Computational Model of the Electrically or Acoustically Evoked Compound Action Potential in Cochlear Implant Users with Residual Hearing

Abstract

Objective: In cochlear implant users with residual acoustic hearing, compound action potentials (CAPs) can be evoked by acoustic (aCAP) or electric (eCAP) stimulation and recorded through the electrodes of the implant. We propose a novel computational model to simulate aCAPs and eCAPs in humans, considering the interaction between combined electric-acoustic stimulation that occurs in the auditory nerve. Methods: The model consists of three components: a 3D finite element method model of an implanted cochlea, a phenomenological single-neuron spiking model for electric-acoustic stimulation, and a physiological multi-compartment neuron model to simulate the individual nerve fiber contributions to the CAP. Results: The CAP morphologies closely resembled those known from humans. The spread of excitation derived from eCAPs by varying the recording electrode along the cochlear implant electrode array was consistent with published human data. The predicted CAP amplitude growth functions largely resembled human data, with deviations in absolute CAP amplitudes for acoustic stimulation. The model reproduced the suppression of eCAPs by simultaneously presented acoustic tone bursts for different masker frequencies and probe stimulation electrodes. Conclusion: The proposed model can simulate CAP responses to electric, acoustic, or combined electric-acoustic stimulation. It considers the dependence on stimulation and recording sites in the cochlea, as well as the interaction between electric and acoustic stimulation in the auditory nerve. Significance: The model enhances comprehension of CAPs and peripheral electric-acoustic interaction. It can be used in the future to investigate objective methods, such as hearing threshold assessment or estimation of neural health through aCAPs or eCAPs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Daniel Kipping, Yixuan Zhang, Waldo Nogueira. 2024-04-18. A Computational Model of the Electrically or Acoustically Evoked Compound Action Potential in Cochlear Implant Users with Residual Hearing. https://doi.org/10.1109/tbme.2024.3410686

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Lens-Aware Differentiable Beamforming for In Vivo Distributed Aberration Correction with Curvilinear Transducers

We previously introduced ultrasound autofocusing, an iterative model-based aberration correction technique that estimates local sound speed and incorporates it into beamforming to correct image distortion from heterogeneous media. In this work, we extend ultrasound autofocusing to curvilinear arrays and introduce advancements to the underlying model. A differentiable bent-ray tracing approach accounts for refraction through the transducer lens, while a new adaptive-grid initialization accounts for changes in speckle position with sound speed. The method is validated in silico and in calibrated sound speed phantoms. Our distributed aberration-correction method is then applied to a first large-scale in vivo evaluation comprising 313 liver acquisitions from 76 high-BMI human subjects. In images containing anechoic regions, contrast and CNR improved by $1.38 \pm 1.60$ dB (+18.0%) and $0.09 \pm 0.14$ (+10.2%), respectively. Improvements were also observed in speckle brightness (+20.3%), coherence factor (+13.1%), lag-one coherence (+2.7%), common-midpoint correlation coefficient (+0.7%), and common-midpoint phase error (-9.1%; lower is better), with all metric improvements statistically significant. Target structure and visibility also improved significantly. These results demonstrate the potential of ultrasound autofocusing for clinically applicable distributed aberration correction.

physics.med-ph

Closing the gap: A two-site MLC matching study for Varian TrueBeam Linear Accelerators

Purpose: To characterise and harmonise multileaf collimator (MLC) gap calibrations across four beam-matched TrueBeams and assess a common planning-system model. Methods: Four Millennium 120 TrueBeams were studied: TS1 (installed 2016), TS5 (2021), TS6 and TS7 (2024). Gap calibration was assessed by RayStation sweeping-gap measurement, feeler gauge, the EPID-based Stakitt Fence test, and a leaf-resolved Machine Performance Check (MPC). TS1 and TS5 were adjusted in service mode, TS5 also for MLC centreline offset. Ten clinical VMAT plans, and six acquired earlier on an unadjusted reference machine, were measured with an ArcCHECK 1220 against doses computed at 0.053 and 0.065 cm x-offsets in RayStation. Results: The 2024 machines had wider baseline gaps. All four methods registered the TS1 and TS5 changes in the same direction and within 0.060 mm of predictions. After adjustment TS1 and TS5 lay within the TS6/TS7 sweeping-gap and Stakitt ranges but read wider on the feeler gauge. A conventional picket-fence test on TS5 registered no change. Raising the RayStation x-offset parameter increased mean local 3%/2 mm pass rates from 92.3% to 94.8% on TS1 and 93.2% to 95.6% on TS5; all 24 plan-criterion pairs improved on the unadjusted machine. Conclusions: Mechanical, EPID, and sweeping gap MLC gap measurements responded consistently to the MLC adjustment, moving TS1 and TS5 towards the local reference range. Every global 3%/2 mm ArcCHECK measurement exceeded the TG-218 95% limit.

physics.med-ph

The Magnetic-Resonance-Based, Omni-Directional Hydrophone

This study aims to improve the directional sensitivity of magnetic resonance hydrophone (MRH) methods and allow for detection and quantification of waves traveling at any angle relative to the MRI bore-axis. Doing so may better integrate MRH technology into transcranial focused ultrasound neuromodulation therapies. Existing MRH implementations only encode acoustic waves propagating along the MRI bore-axis, precluding use with wide aperture transducers or temporal skull windows. This study expands the MRH technique into a MR-based Omni-directional Hydrophone (MRbOH) that can quantify the magnitude pressure and phase of waves propagating at any angle. The MRbOH method formulates a system of equations around uniquely sensitized images that, when solved, introduce sensitivity to oblique acoustic waves. The method was tested experimentally in a gel phantom that was insonated at $0^o$, $45^o$, and $90^o$ relative to the MRI bore-axis. The method reconstructed pressure maps at all three angles of propagation, including features such as peak pressure, partial reflections, and beamwidth. The MRbOH method measured an average beamwidth of 16.2 mm while the hydrophone measured a beamwidth of 14.5 mm. Average hydrophone-measured pressure within a region of interest was 416.19 $\boldsymbol{\pm}$ 78.07 kPa. Average MRbOH measured pressure in the same region at $0^o$, $45^o$, and $90^o$, respectively, were: 285.5 $\boldsymbol{\pm}$ 49.5 kPa, 193.13 $\boldsymbol{\pm}$ 43.55 kPa, 194.7 $\boldsymbol{\pm}$ 47.38 kPa. In addition, the study presents and experimentally verifies a statistical noise model that predicted the mean with $R^2 = 0.85$, and standard deviation with $R^2 = 0.83$.

physics.med-ph