Search arXivSearch

arXiv · 2406.17143

A Hardware/Firmware-Based Switching Gate Multiplexing Method for Pulse Mode Radiation Detectors

Abstract

We present a hardware/firmware-based switching gate multiplexing method for pulse mode radiation detectors that can combine many detector signals into two readout channels. One readout channel passes the signal of the multiplexed detector that "fired" first, and the other channel provides a variable-width logic pulse, i.e., a pulse width modulation (PWM) signal, that identifies the active detector. The multiplexed output pulse is produced by passing the first active detector's signal to a fan-in circuit by gating on the corresponding channel for a fixed duration while blocking all other detector signals. It does this using individual analog switches for all the detector signals. Each switch is controlled by a fixed width logic pulse that is triggered by the arrival of the first active detector pulse. Both the fixed width logic pulse and the PWM signal are generated using a field-programmable gate array (FPGA). To demonstrate the proposed multiplexing method, a prototype four-channel multiplexer was developed for use with four NaI(Tl) detectors. The performance of the multiplexer was evaluated in terms of its ability to retain energy resolution, timing resolution, and original pulse shape. The proposed multiplexing method showed very little degradation in energy resolution and timing resolution or alteration of pulse shape. The switching gate feature of the proposed method enables the multiplexer output to have very low noise contribution from the inactive channels. This multiplexing technique also has the unique capability of isolating and recovering the first active detector's output pulse in cases where there is overlap between pulses from different detectors in a single digitized record. These features make the proposed hardware/firmware-based switching gate multiplexing method very promising for application to large radiation detector networks.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Md Faisal Rahman, John Mattingly. 2024-06-24. A Hardware/Firmware-Based Switching Gate Multiplexing Method for Pulse Mode Radiation Detectors. https://doi.org/10.1016/j.nima.2024.169574

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det