Search arXivSearch

arXiv · 2412.01515

Performance of the LABDOS01 spectrometer in dosimetric measurements

Abstract

This paper describes the performance of the LABDOS01, a silicon diode-based spectrometer suitable for dose measurements in mixed radiation fields. The instrument is currently being used in two high-altitude environmental dose monitoring projects: SAMADHA (South Atlantic Magnetic Anomaly Dosimetry at High Altitude) at Chacaltaya (Bolivia, 5240 m a.s.l.) and CORDIAL (COsmic Rays Dosimetry In Antarctic Latitudes) at the Concordia station (Antarctica, 3233 m a.s.l.). Before installing two of these devices at the measurement sites, the detectors were tested on flight routes covering a wide range of geomagnetic latitudes. The collected dosimetric data were compared with the expectations derived by the CARI-7A software, which provides the absorbed dose rate in silicon due to cosmic ray secondaries at a given position on the Earth. The measured dose rates along the flights at variable altitude and rigidity cutoff agree well with the simulated ones. By analyzing the spectrum of the energy deposited in the silicon layer, we derive an empirical method to approximately evaluate the ambient dose equivalent $H^{*}(10)$, a quantity directly related to the biological damage caused by environmental radiation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Silvia Vernetto, Carlo Francesco Vigorito, Martin Kakona, Roman Dvorak, Satoshi Kodaira, Stefano Cirilli, Fabio Romanelli, Daniele Savietto, Alba Zanini. 2024-12-02. Performance of the LABDOS01 spectrometer in dosimetric measurements. https://arxiv.org/abs/2412.01515

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Mortality of ultra-thin LGADs and PiN diodes from high energy deposition

Low Gain Avalanche Diodes are prime candidates for high-resolution timing applications in High Energy Physics, Nuclear science, and several other fields. Operating these devices in high-radiation environments presents various hazards, including the risk of their permanent degradation or destruction caused by effects such as Single Event Burnout. Studies using minimum ionizing particles found a greatly reduced Single Event Burnout risk by operating below a bias voltage corresponding to an average electric field of 12 V/$μ$m - however, as high energy particle colliders produce a wide energy spectrum of radiation, it is crucial to understand this phenomenon and other possible damage mechanisms at energy deposition levels greater than those of minimum ionizing particles. This was achieved by pre-irradiating LGADs and PiN diodes with active thicknesses of 20, 30, and 50 $μ$m up to 1.5 $\times$ 10$^{15}$ $\mathrm{n_{eq}/cm^2}$, and exposing them to beams of protons and heavy ions (C, O, Fe, Au) at the BNL Tandem van de Graaff accelerator. Several mortality categories were observed, defined by different electrical and mechanical damage signatures. This furthers our understanding of permanent radiation damage of silicon devices, crucial towards mitigating Single Event Burnout and other damage mechanisms to safely operate future detectors.

physics.ins-det

Small but Tubby: A Magnetic Loop Antenna Made from 100 mm Copper Tubing

This paper presents the electrical model, key equations, and practical construction of a small transmitting magnetic loop antenna built from unusually large 100 mm diameter copper tubing. The large conductor surface area and wide-area transitions to the vacuum capacitors were designed to minimize resistive losses. The frequency range from 1.8 MHz to 31 MHz is unusually wide. Frequency, impedance matching, and azimuth are all adjusted automatically by servo motors. Indoor losses originate predominantly from near-field coupling to the environment rather than from the antenna itself. Temperature-rise measurements confirm that the bulk of the dissipated power is absorbed by the environment, not by the antenna components. The conducted H-field measurements demonstrate good agreement between the measured H-field and the theoretical free-space H-field calculated from the antenna geometry and an estimated loop current. The loop current was estimated from the measured antenna bandwidth and the applied transmit power. The antenna was developed for indoor operation where outdoor installation is not possible.

physics.ins-det

Construction and characterisation of the DarkSide-20k veto silicon photo-multiplier tiles

Silicon photo-multipliers (SiPMs) are state-of-the-art sensors capable of detecting a single photoelectron under cryogenic conditions, with potentially lower radioactivity than widely used photomultiplier tubes. The DarkSide-20k experiment, designed to perform direct dark matter searches using liquid argon as the target material, employs SiPM technology to detect interactions in the active detector volumes, including the central dual-phase Time Projection Chamber and the Inner and Outer Veto volumes. The vetoes are designed to discriminate against radiogenic neutron and cosmic muon backgrounds associated with the dark matter search. This paper describes the completed production and test protocols for the "Veto Tiles" (called vTiles, arrays of 24 SiPMs integrated on a printed circuit board providing the power distribution and signal amplification); 16 vTiles are grouped into "Veto Photo-Detector Units" to instrument the Inner Veto volume. Each vTile underwent detailed testing at room and cryogenic temperatures, confirming stable operation, high signal-to-noise ratio, and low radioactive contamination, demonstrating the robustness of the proposed design for cryogenic conditions. The final production yield exceeded 87%, surpassing the 80% requirement and corresponding to 1920 Veto Tiles to populate 120 Veto Photo-Detector Units, plus an additional 6% as spares.

physics.ins-det