Myocardial T1 mapping at 5T using multi-inversion recovery real-time spoiled GRE

Abstract

Objective: To develop an accurate myocardial T1 mapping technique at 5T using Look-Locker-based multiple inversion-recovery with the real-time spoiled gradient echo (GRE) acquisition. Approach: The proposed T1 mapping technique (mIR-rt) samples the recovery of inverted magnetization using the real-time GRE and the images captured during diastole are selected for T1 fitting. Multiple-inversion recoveries are employed to increase the sample size for accurate fitting. The T1 mapping method was validated using Bloch simulation, phantom studies, and in 16 healthy volunteers at 5T. Main results: In both simulation and phantom studies, the T1 values measured by mIR-rt closely approximate the reference T1 values, with errors less than 3%, while the conventional MOLLI sequence underestimates T1 values. The myocardial T1 values at 5T are 1553 +/- 52 ms, 1531 +/- 53 ms, and 1526 +/- 60 ms (mean +/- standard deviation) at the apex, middle, and base, respectively. The T1 values measured by MOLLI (1350 +/- 48 ms, 1349 +/- 47 ms, and 1354 +/- 45 ms at the apex, middle, and base) were significantly lower than those of mIR-rt with p<0.001 for all three layers. The mIR-rt sequence method used in our study provides high reproducibility, particularly in the middle slices, supporting its practical relevance for myocardial T1 mapping. Significance: The proposed method is feasible for myocardial T1 mapping at 5T and provides better accuracy than the conventional MOLLI sequence.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Linqi Ge, Yihang Zhang, Huibin Zhu, Lang Zhang, Yihang Zhou, Haifeng Wang, Dong Liang, Hairong Zheng, Yanjie Zhu. 2025-08-28. Myocardial T1 mapping at 5T using multi-inversion recovery real-time spoiled GRE. https://arxiv.org/abs/2501.07081

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Voxel-Matching NORDIC: Non-local patch formation by time-series similarity increases tSNR in high-resolution BOLD fMRI

Submillimeter functional magnetic resonance imaging (fMRI) based on blood-oxygenation-level-dependent (BOLD) signal enables the study of brain function at the submillimeter level, uncovering insights into fine-scale organisations like cortical layers and columns. However, its inherently low contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) often limit its reliability and applicability. Noise Reduction with Distribution Corrected Principal Components Analysis (NORDIC PCA) is a locally low-rank denoising algorithm that reduces thermal noise levels in BOLD fMRI in a local patch manner. However, local patches often contain a mixture of signals from multiple tissues that negatively affects the low-rank structure of the patches, which limits the denoising capabilities of the algorithm. We propose an alternative approach for patch formation by gathering similar non-local voxels, dubbed voxel-matching (VM) NORDIC. The results on submillimeter-resolution BOLD fMRI data indicate that VM-NORDIC effectively promotes the low rank of the patches by boosting signal redundancy, allowing for more efficient noise attenuation. Moreover, the method barely affects spatial smoothness due to the non-local voxel selection based on time-series similarity. In particular, VM-NORDIC outperforms standard NORDIC with default local patching (Standard-NORDIC) in terms of temporal SNR (tSNR) (~9-90% larger than Standard-NORDIC; ~23-250% larger than the original) and spatial smoothness estimates (~20% of the smoothness induced by Standard-NORDIC). These improvements are fundamental to improving the validity and precision of fMRI studies at submillimeter resolutions.

physics.med-ph

Computed Tomography Reconstruction Algorithm Using Markov Random Field Model

X-ray computed tomography (CT) reveals the materials' internal structures non-destructively from a tilt series of projected images. Filtered back projection (FBP) is a widely-adopted reconstruction algorithm in CT owing to its small computational cost. Under low-dose or sparse-view conditions, however, FBP often amplifies noise, severely degrading the reconstructed images. In this study, we evaluated the performance of a Bayesian CT reconstruction algorithm based on the Markov random field model under such adverse conditions. Through simulations, we demonstrated that the proposed algorithm shows higher reconstruction performance than FBP under both low-dose and sparse-view conditions. The hyperparameters are estimated by minimizing the Bayesian free energy, enabling adaptive reconstruction that reflects the noise characteristics of the observed projection data. These results suggest that the proposed algorithm can broaden the applicability of CT to dose-sensitive applications and time-constrained measurements, where only limited observed projection data are available.

physics.med-ph

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion--ion recombination and self-consistent space-charge electric field calculations. The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions. The implementation was validated against analytical and numerical models from the literature. Excellent agreement was obtained for the free electron fraction (FEF), CCE, induced current, and electric field evolution under UHDR conditions. The simulations show that space charge can locally increase the electric field by more than a factor of four or reduce it to nearly zero in some regions of the chamber. The results further suggest that the reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment. This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.

physics.med-ph