Underestimation of systolic pressure in cuff-based blood pressure measurement

Abstract

High blood pressure (hypertension) is the number one risk factor for premature death. Hypertension is asymptomatic, so blood pressure must be regularly monitored to diagnose it. In auscultatory blood pressure measurement, a patient's systolic (maximum) and diastolic (minimum) blood pressure are inferred from the pressure in an inflatable cuff wrapped around the arm. This technique is the gold standard against which all other non-invasive devices are validated. However, auscultatory measurements systematically underestimate systolic blood pressure and overestimate diastolic blood pressure. Overestimation is attributed to the increased cuff pressure needed to occlude the artery because of the surrounding tissue and arterial stiffness. In contrast, the cause of systolic underestimation, which leads to potentially a third of systolic hypertension cases being missed, has remained unclear. When the cuff is inflated beyond the systolic blood pressure, the blood flow to the vessels downstream of the cuff is cut off. The pressure in these downstream vessels drops to a low plateau. We have developed a novel experimental rig that shows that the low downstream pressure is the key cause of the underestimation of systolic blood pressure. The lower the downstream pressure, the greater the underestimation. Our results yield a simple physical model for the underestimation of systolic pressure in our rig and in the human body. Understanding the physics behind the underestimation of systolic blood pressure paves the way for developing strategies to mitigate this error.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Kate Bassil, Anurag Agarwal. 2025-01-17. Underestimation of systolic pressure in cuff-based blood pressure measurement. https://arxiv.org/abs/2501.10232

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Fingertip Micro-Motion as a Source of Respiratory Information During Sleep Using Triaxial Accelerometers

Objective: Triaxial accelerometers (TAAs) are widely used in home care medicine. This study investigates whether TAA signals recorded at the fingertip encode respiratory information, particularly instantaneous respiratory rate (IRR) and respiratory effort, during sleep. Method: We propose an antiderivative-based nonlinear transformation to convert TAA signals into a respiratory surrogate, termed TAA-resp. To quantify the embedded respiration-induced oscillation, a modern time-frequency analysis tool is applied to derive an index, referred to as the {\em respiratory motion index} (RMI). The proposed TAA-resp and RMI are validated on two datasets comprising 39 Asians and 10 white or African Americans with full-night recordings from simultaneous polysomnography (PSG) and a fingertip TAA measurement. Criteria for labeling TAA-resp signal quality as good, moderate, or poor are established, and two experts' independent annotations with consensus are obtained. Result: On average, TAA-resp encodes high-quality respiratory information in over 21.25%$\pm$16.11% of full-night recordings, reaching up to 58.33% in some cases. TAA-resp shows stronger correlation with thoracic and abdominal motion than with airflow, indicating predominant capture of respiratory effort. High-quality TAA-resp segments offer an accurate IRR estimate with root mean square error $0.027\pm 0.022$ Hz. RMI is higher for high-quality segments and lower for poor-quality segments, and its distribution aligns with physiology, with higher values during REM, N2, and N3 sleep and in the absence of apnea or hypopnea events. In leave-one-subject-out cross-validation, RMI predicts quality labels with 0.84 sensitivity and 0.88 specificity. Conclusion: Fingertip-mounted TAAs encode meaningful respiratory information intermittently, and our method recycles underutilized respiratory information.

physics.med-ph↗

Characterization and Mitigation of Boundary Sampling Artifacts in Electron Phase Space Generation for Clinical FLASH-RT Deployment

Applicator-specific phase space (PHSP) files recorded at the aperture exit reduce Monte Carlo dose calculation time by 30-50% for electron FLASH radiotherapy. However, positioning PHSP scoring planes coincident with the applicator-air interface introduces boundary sampling artifacts. This study characterizes these artifacts in Geant4-based simulations and demonstrates their mitigation. PHSP files were generated using GAMOS 6.2.0 for a 9 MeV Mobetron UHDR model across twelve clinical aperture configurations (2.5-10 cm diameter). Three scoring plane positions were evaluated relative to the physical aperture exit: coincident with the interface, 0.1 mm downstream, and 1 mm downstream. Scoring at the exact interface produced proximal R50 shifts of up to 2.2 mm and Distance-to-Agreement (DTA) values of 4-6 mm, exceeding clinical acceptance criteria. Artifact severity scaled inversely with aperture diameter, with the smallest configurations most severely affected. A 0.1 mm offset partially restored the primary electron energy spectrum but failed to recover the bremsstrahlung tail. A 1 mm offset fully resolved all artifacts, achieving mean DTA values within 2.0 mm, equivalent to or better than linac-exit references. These artifacts arise from the degenerate behavior of the fUseSafety step-limitation algorithm when safety equals zero at exact material boundaries, producing incomplete secondary electron equilibration and suppressed bremsstrahlung production. Angular distribution analysis revealed a near-forward particle pileup in the 0 mm PHSP and a deficit of large-angle secondaries recovered by the 1 mm offset. A 1 mm downstream offset fully mitigates these artifacts while introducing negligible perturbation to primary beam characteristics. This requirement applies to any Geant4-based framework (including TOPAS and GATE) scoring PHSP files at material exit surfaces.

physics.med-ph↗

Melanin- and linewidth-corrected pulse-oximetry

Current pulse oximeters tend to overestimate arterial saturation in darkly pigmented patients, increasing occult hypoxemia rates. One proposed mechanism is spectral coloring: melanin contributes to reshaping the detected spectrum of a broad-linewidth source, making the wavelength-dependent parameters behind calibration skin-tone-dependent. Diffuse optics theory allows for a melanin- and linewidth-corrected calibration equation, which to our knowledge has never been tested in vivo. In this work, we fit the empirical linear equation, the diffuse optics theoretical equation, and its corrected form to the public OpenOximetry Repository---considering 98 patients with paired arterial blood-draws, raw photoplethysmograms, and skin reflectance spectra. Accuracy is practically indistinguishable between the three approaches, yet the saturation error's dependence on melanin falls from +2.8 to +1.2 to +0.1 percentage points per unit melanin volume fraction---consistent with the spectral-coloring theory. None of these dependencies on melanin are statistically resolved here---a cohort roughly ten times larger is needed---so the correction is backed by theory and its expected direction. All three calibration approaches carry two degrees of freedom: the theoretical equation drops into existing calibration procedures, and its melanin-aware form needs a per-patient skin-tone measurement. This work demonstrates how the theoretical pulse oximetry model can be applied in calibration.

physics.med-ph↗