Search arXivSearch

arXiv · 2502.07630

Rethinking Timing Residuals: Advancing PET Detectors with Explicit TOF Corrections

Abstract

PET is a functional imaging method that visualizes metabolic processes. TOF information can be derived from coincident detector signals and incorporated into image reconstruction to enhance the SNR. PET detectors are typically assessed by their CTR, but timing performance is degraded by various factors. Research on timing calibration seeks to mitigate these degradations and restore accurate timing information. While many calibration methods use analytical approaches, machine learning techniques have recently gained attention due to their flexibility. We developed a residual physics-based calibration approach that combines prior domain knowledge with the power of machine learning models. This approach begins with an initial analytical calibration addressing first-order skews. The remaining deviations, regarded as residual effects, are used to train machine learning models to eliminate higher-order skews. The key advantage is that the experimenter guides the learning process through the definition of timing residuals. In earlier studies, we developed models that directly predicted the expected time difference, which offered corrections only implicitly (implicit correction models). In this study, we introduce a new definition for timing residuals, enabling us to train models that directly predict correction values (explicit correction models). The explicit correction approach significantly simplifies data acquisition, improves linearity, and enhances timing performance from $371 \pm 6$ ps to $281 \pm 5$ ps for coincidences from 430 keV to 590 keV. Additionally, the new definition reduces model size, making it suitable for high-throughput applications like PET scanners. Experiments were conducted using two detector stacks composed of $4 \times 4$ LYSO:Ce,Ca crystals ($3.8\times 3.8\times 20$ mm$^{3}$) coupled to $4 \times 4$ Broadcom NUV-MT SiPMs and digitized with the TOFPET2 ASIC.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Stephan Naunheim, Luis Lopes de Paiva, Vanessa Nadig, Yannick Kuhl, Stefan Gundacker, Florian Mueller, Volkmar Schulz. 2025-04-22. Rethinking Timing Residuals: Advancing PET Detectors with Explicit TOF Corrections. https://doi.org/10.3389/fphy.2025.1570925

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det