Accurate myocardial T1 mapping at 5T using an improved MOLLI method: A validation study

Abstract

Background: Accurate myocardial T1 mapping at 5T remains a technical challenge due to field inhomogeneity and prolonged T1 values. The aim of this study is to develop an accurate and clinically applicable myocardial T1 mapping technique for 5T magnetic resonance imaging (MRI) systems and validate its performance in a multicenter study. Methods: The proposed method is based on a 5-(3)-3 MOLLI sequence, dubbed combined-correction MOLLI (coMOLLI), which corrects for both inversion efficiency and readout-induced signal disturbances. Specifically, coMOLLI employs a gradient recalled echo (GRE) readout rather than the commonly used bSSFP readout. Its signal evolution is modeled to estimate T1 values, which incorporates both inversion efficiency and readout disturbances to improve fitting accuracy. To further enhance accuracy, the inversion pulse was redesigned under hardware constraints and the observed B0 and B1 variations over the heart at 5T, using adiabatic hyperbolic secant (HSn) and tangent/hyperbolic tangent (Tan/Tanh) pulses. The method was validated in phantom experiments, as well as in 21 healthy volunteers and 9 patients. Results: The optimized inversion pulse at 5T is the Tan/Tanh pulse with A = 10 kHz, K_s = 4, k = 22, and T_p = 8 ms. In phantom studies, coMOLLI showed high accuracy versus reference IR-FSE, yielding relative errors within 5% for all nine vials. In vivo studies, the average native myocardial T1 values across 21 healthy volunteers were 1468 +/- 48 ms, 1514 +/- 39 ms, and 1545 +/- 50 ms, and blood T1 values were 2182 +/- 132 ms, 2124 +/- 153 ms, and 2131 +/- 158 ms for apical, middle, and base slices, respectively. Conclusion: The coMOLLI method demonstrated high accuracy in phantom studies and feasibility in vivo studies. By adopting the widely used 5-(3)-3 MOLLI acquisition scheme, it shows potential for clinical cardiac imaging at 5T.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Linqi Ge, Yinuo Zhao, Yubo Guo, Yuanyuan Liu, Yihang Zhou, Haifeng Wang, Dong Liang, Hairong Zheng, Yining Wang, Yanjie Zhu. 2025-10-15. Accurate myocardial T1 mapping at 5T using an improved MOLLI method: A validation study. https://arxiv.org/abs/2503.08425

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Voxel-Matching NORDIC: Non-local patch formation by time-series similarity increases tSNR in high-resolution BOLD fMRI

Submillimeter functional magnetic resonance imaging (fMRI) based on blood-oxygenation-level-dependent (BOLD) signal enables the study of brain function at the submillimeter level, uncovering insights into fine-scale organisations like cortical layers and columns. However, its inherently low contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) often limit its reliability and applicability. Noise Reduction with Distribution Corrected Principal Components Analysis (NORDIC PCA) is a locally low-rank denoising algorithm that reduces thermal noise levels in BOLD fMRI in a local patch manner. However, local patches often contain a mixture of signals from multiple tissues that negatively affects the low-rank structure of the patches, which limits the denoising capabilities of the algorithm. We propose an alternative approach for patch formation by gathering similar non-local voxels, dubbed voxel-matching (VM) NORDIC. The results on submillimeter-resolution BOLD fMRI data indicate that VM-NORDIC effectively promotes the low rank of the patches by boosting signal redundancy, allowing for more efficient noise attenuation. Moreover, the method barely affects spatial smoothness due to the non-local voxel selection based on time-series similarity. In particular, VM-NORDIC outperforms standard NORDIC with default local patching (Standard-NORDIC) in terms of temporal SNR (tSNR) (~9-90% larger than Standard-NORDIC; ~23-250% larger than the original) and spatial smoothness estimates (~20% of the smoothness induced by Standard-NORDIC). These improvements are fundamental to improving the validity and precision of fMRI studies at submillimeter resolutions.

physics.med-ph

Computed Tomography Reconstruction Algorithm Using Markov Random Field Model

X-ray computed tomography (CT) reveals the materials' internal structures non-destructively from a tilt series of projected images. Filtered back projection (FBP) is a widely-adopted reconstruction algorithm in CT owing to its small computational cost. Under low-dose or sparse-view conditions, however, FBP often amplifies noise, severely degrading the reconstructed images. In this study, we evaluated the performance of a Bayesian CT reconstruction algorithm based on the Markov random field model under such adverse conditions. Through simulations, we demonstrated that the proposed algorithm shows higher reconstruction performance than FBP under both low-dose and sparse-view conditions. The hyperparameters are estimated by minimizing the Bayesian free energy, enabling adaptive reconstruction that reflects the noise characteristics of the observed projection data. These results suggest that the proposed algorithm can broaden the applicability of CT to dose-sensitive applications and time-constrained measurements, where only limited observed projection data are available.

physics.med-ph

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates: A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination

Ultra-high dose rate (UHDR) irradiation used in FLASH radiotherapy can lead to significant reductions in charge collection efficiency (CCE) in plane-parallel ionisation chambers (PPICs) due to enhanced ion recombination, while the high charge densities involved can also induce strong space-charge-induced electric-field distortions. To investigate these coupled effects, we extended the Garfield++ Monte Carlo particle-transport framework by implementing ion--ion recombination and self-consistent space-charge electric field calculations. The developed Monte Carlo model couples particle transport, electron attachment, recombination processes, and dynamic electric-field distortions. The implementation was validated against analytical and numerical models from the literature. Excellent agreement was obtained for the free electron fraction (FEF), CCE, induced current, and electric field evolution under UHDR conditions. The simulations show that space charge can locally increase the electric field by more than a factor of four or reduce it to nearly zero in some regions of the chamber. The results further suggest that the reduction of the CCE under UHDR conditions is mainly driven by the decrease of the FEF caused by electric-field-dependent electron attachment. This finding indicates that the complex problem of recombination under UHDR conditions may be largely governed by the evolution of the FEF, opening promising perspectives for the development of improved analytical models and real-time correction methods for ionisation chamber dosimetry under UHDR irradiation conditions. This work provides a flexible and self-consistent Monte Carlo framework for investigating recombination phenomena and improving dosimetry models for FLASH radiotherapy.

physics.med-ph