Invasion depth estimation of gastric cancer in early stage using circularly polarized light scattering: Phantom studies

Abstract

Depolarization of circularly polarized light due to multiple scattering in turbid media provides size distributions of scatterer. Applied it to the biological tissues as turbid media, scatters correspond to cell nuclei, which is abnormally grown in cancerous tissues. Therefore, the invasion depth of early-staged cancer can be estimated by comparisons of circular polarization of scattered light. In this study, we fabricated the optical phantoms made of resin and polystyrene beads to verify this technique by systematic experiments. The single-layered phantoms containing only one size of beads exhibits uniform monodispersed scattering media. Polarization images taken by a circular polarization imaging camera show systematically changes to the numerical density of scatters. Healthy and cancerous tissue phantoms exhibiting the lengths of mean free paths close to actual biological tissues were stacked to be bi-layered phantoms which imitates early-staged cancers. The averaged circular polarization values obtained from the images captured with the polarization camera are indicative of obvious changes depending on the thickness of cancerous layer of phantoms.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Mike R. Maskey, Mitsuo Fukuno, Akane Saito, Asato Esumi, Takahiro Kuchimaru, Nozomi Nishizawa. 2025-04-24. Invasion depth estimation of gastric cancer in early stage using circularly polarized light scattering: Phantom studies. https://arxiv.org/abs/2504.17161

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Explicit pixel-value correction of lateral response artifacts in radiochromic film dosimetry: Evaluation of single-, double-, and triple-channel methods

Background:Flatbed scanners introduce a lateral response artifact (LRA) perpendicular to lamp motion. Because LRA varies with position, pixel value (PV), color channel, and film model, it may violate the channel-common perturbation assumption of multichannel dosimetry. Purpose:To characterize PV-domain LRA in EBT4 and EBT-XD films, compare Lewis and Full corrections, and evaluate their effects on single-, double-, and triple-channel dosimetry. Methods:Seven films were scanned at seven lateral positions using a cyclic-shift design. For each position and RGB channel, center-equivalent response was modeled as PVc = A + B * PV. Lewis correction used two films with linear interpolation of A and B; Full correction used all center-local pairs with PCHIP interpolation of PVc - PV. Performance was assessed by portrait-landscape profile consistency and dose accuracy. Results: LRA was affine in PV, asymmetric about scanner center, and dependent on film model and channel. Both corrections reduced profile differences. Mean absolute profile differences across MU levels ranged from 1.4%-5.3% for EBT4 and 2.4%-6.8% for EBT-XD. The methods performed similarly for EBT4, while Full correction reduced discrepancies for several EBT-XD double- and triple-channel methods. Residual blue-channel errors limited blue, red-blue, and triple-channel performance. At 500 MU and above, red showed the best dose agreement; at 100 MU, both corrections increased deviations for several single-channel estimates. Full correction produced smaller deviations than Lewis in several EBT-XD conditions. Conclusions:LRA should be corrected in the raw-PV domain before dose reconstruction. Full correction better preserved profile consistency and dose accuracy, particularly for EBT-XD. Because improved profile consistency did not always improve central-dose agreement, both should be evaluated separately in commissioning.

physics.med-ph

Integration of Spectral CT with PET and SPECT: Bringing Tissue Composition Information to Molecular Imaging

Molecular imaging has been transformed by integrating positron emission tomography (PET) or single-photon emission computed tomography (SPECT) with x-ray computed tomography (CT). However, x-ray CT in hybrid imaging is used primarily for anatomical localization and for attenuation and scatter correction of emission data. Its ability to characterize tissue composition remains underutilized. Spectral CT, including dual-energy CT (DECT) and photon-counting CT (PCCT), can provide material-specific information, such as images of iodine concentration, bone or calcium fractions, electron-density, effective atomic number and other spatially-varying material properties. These capabilities create an opportunity to bring tissue-composition information into molecular imaging. In addition, spectral CT can also reduce CT artifacts and provide more accurate corrections for emission data. In this review, we discuss the technical basis, integration pathways, and translational opportunities for combining spectral CT with molecular imaging. We review sequential and integrated PET/DECT approaches, emerging PET/PCCT concepts, PET-enabled spectral CT, and extensions to SPECT/CT. We then examine how spectral CT information may improve attenuation correction, scatter correction, tissue-fraction correction, positron range correction, reconstruction priors, and dosimetry. Clinical and translational applications include contrast-enhanced molecular imaging, vascular and perfusion imaging, quantitative bone marrow imaging, musculoskeletal imaging, and theranostics. Overall, spectral CT integration may move hybrid molecular imaging toward tissue-composition-informed molecular imaging, in which radiotracer signals are interpreted within their material context.

physics.med-ph

Non-circular scan trajectories for reducing cone-beam artifacts in Gamma Knife CBCT images: a simulation study

Objective. Gamma Knife cone-beam computed tomography (CBCT) images are deteriorated by cone-beam artifacts whose magnitude increases along the superior direction. In this study, a novel scan trajectory compatible with the Gamma Knife CBCT system is optimized to reduce cone-beam artifacts, with the aim to replace the current 200-degree single-arc scan. Approach. Data sampling analysis with tomographic incompleteness maps indicates the level of undersampling across the field of view based on the geometry of the system and of a scan trajectory. Moreover, CBCT simulations are performed from a virtual phantom with disks aligned along the axial direction and from a CT reconstruction of a stereotactic end-to-end validation (STEEV) phantom. CBCT projections are simulated for a given scan trajectory through a polychromatic forward projection model with added noise and scatter, then the CBCT image is reconstructed using an iterative algorithm which minimizes weighted least squares. Main results. Both the incompleteness analysis and the CBCT simulations indicate adding lines to the current single-arc trajectory is more efficient to reduce cone-beam artifacts than adding more arcs, both in terms of number of additional projections and new artifacts. A line-arc-line trajectory with source axial steps of 3.5 mm removes virtually all cone-beam artifacts. The widths of the cone-beam artifacts created by the disks show a positive correlation between the artifact magnitude and the incompleteness value. Significance. A line-arc-line scan trajectory is promising to reduce cone-beam artifacts of the Gamma Knife CBCT images while being a compatible and reasonable trajectory for the current system design.

physics.med-ph