Bimodal Cochlear Implants: Measurement of the Localization Performance as a Function of Device Latency Difference

Abstract

Bimodal cochlear implant users show poor localization performance. One reason for this is a difference in the processing latency between the hearing aid and the cochlear implant side. It has been shown that reducing this latency difference acutely improves the localization performance of bimodal cochlear implant users. However, due to the frequency dependency of both the device latencies and the acoustic hearing ear, current frequency-independent latency adjustments cannot fully compensate for the differences, leaving open which latency adjustment is best. We therefore measured the localization performance of eleven bimodal cochlear implant users for multiple cochlear implant latencies. We confirm previous studies that adjusting the interaural latency improves localization in all eleven bimodal cochlear implant users. However, the latency that leads to the best localization performance for most subjects was not at the latency calculated to compensate for the interaural difference at intermediate frequencies (1 kHz). Nine of eleven subjects localized best with a cochlear implant latency that was slightly shorter than the calculated latency compensation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Rebecca C. Felsheim, Sabine Hochmuth, Alina Kleinow, Andreas Radeloff, Mathias Dietz. 2025-05-05. Bimodal Cochlear Implants: Measurement of the Localization Performance as a Function of Device Latency Difference. https://arxiv.org/abs/2505.02555

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Fingertip Micro-Motion as a Source of Respiratory Information During Sleep Using Triaxial Accelerometers

Objective: Triaxial accelerometers (TAAs) are widely used in home care medicine. This study investigates whether TAA signals recorded at the fingertip encode respiratory information, particularly instantaneous respiratory rate (IRR) and respiratory effort, during sleep. Method: We propose an antiderivative-based nonlinear transformation to convert TAA signals into a respiratory surrogate, termed TAA-resp. To quantify the embedded respiration-induced oscillation, a modern time-frequency analysis tool is applied to derive an index, referred to as the {\em respiratory motion index} (RMI). The proposed TAA-resp and RMI are validated on two datasets comprising 39 Asians and 10 white or African Americans with full-night recordings from simultaneous polysomnography (PSG) and a fingertip TAA measurement. Criteria for labeling TAA-resp signal quality as good, moderate, or poor are established, and two experts' independent annotations with consensus are obtained. Result: On average, TAA-resp encodes high-quality respiratory information in over 21.25%$\pm$16.11% of full-night recordings, reaching up to 58.33% in some cases. TAA-resp shows stronger correlation with thoracic and abdominal motion than with airflow, indicating predominant capture of respiratory effort. High-quality TAA-resp segments offer an accurate IRR estimate with root mean square error $0.027\pm 0.022$ Hz. RMI is higher for high-quality segments and lower for poor-quality segments, and its distribution aligns with physiology, with higher values during REM, N2, and N3 sleep and in the absence of apnea or hypopnea events. In leave-one-subject-out cross-validation, RMI predicts quality labels with 0.84 sensitivity and 0.88 specificity. Conclusion: Fingertip-mounted TAAs encode meaningful respiratory information intermittently, and our method recycles underutilized respiratory information.

physics.med-ph↗

Characterization and Mitigation of Boundary Sampling Artifacts in Electron Phase Space Generation for Clinical FLASH-RT Deployment

Applicator-specific phase space (PHSP) files recorded at the aperture exit reduce Monte Carlo dose calculation time by 30-50% for electron FLASH radiotherapy. However, positioning PHSP scoring planes coincident with the applicator-air interface introduces boundary sampling artifacts. This study characterizes these artifacts in Geant4-based simulations and demonstrates their mitigation. PHSP files were generated using GAMOS 6.2.0 for a 9 MeV Mobetron UHDR model across twelve clinical aperture configurations (2.5-10 cm diameter). Three scoring plane positions were evaluated relative to the physical aperture exit: coincident with the interface, 0.1 mm downstream, and 1 mm downstream. Scoring at the exact interface produced proximal R50 shifts of up to 2.2 mm and Distance-to-Agreement (DTA) values of 4-6 mm, exceeding clinical acceptance criteria. Artifact severity scaled inversely with aperture diameter, with the smallest configurations most severely affected. A 0.1 mm offset partially restored the primary electron energy spectrum but failed to recover the bremsstrahlung tail. A 1 mm offset fully resolved all artifacts, achieving mean DTA values within 2.0 mm, equivalent to or better than linac-exit references. These artifacts arise from the degenerate behavior of the fUseSafety step-limitation algorithm when safety equals zero at exact material boundaries, producing incomplete secondary electron equilibration and suppressed bremsstrahlung production. Angular distribution analysis revealed a near-forward particle pileup in the 0 mm PHSP and a deficit of large-angle secondaries recovered by the 1 mm offset. A 1 mm downstream offset fully mitigates these artifacts while introducing negligible perturbation to primary beam characteristics. This requirement applies to any Geant4-based framework (including TOPAS and GATE) scoring PHSP files at material exit surfaces.

physics.med-ph↗

Melanin- and linewidth-corrected pulse-oximetry

Current pulse oximeters tend to overestimate arterial saturation in darkly pigmented patients, increasing occult hypoxemia rates. One proposed mechanism is spectral coloring: melanin contributes to reshaping the detected spectrum of a broad-linewidth source, making the wavelength-dependent parameters behind calibration skin-tone-dependent. Diffuse optics theory allows for a melanin- and linewidth-corrected calibration equation, which to our knowledge has never been tested in vivo. In this work, we fit the empirical linear equation, the diffuse optics theoretical equation, and its corrected form to the public OpenOximetry Repository---considering 98 patients with paired arterial blood-draws, raw photoplethysmograms, and skin reflectance spectra. Accuracy is practically indistinguishable between the three approaches, yet the saturation error's dependence on melanin falls from +2.8 to +1.2 to +0.1 percentage points per unit melanin volume fraction---consistent with the spectral-coloring theory. None of these dependencies on melanin are statistically resolved here---a cohort roughly ten times larger is needed---so the correction is backed by theory and its expected direction. All three calibration approaches carry two degrees of freedom: the theoretical equation drops into existing calibration procedures, and its melanin-aware form needs a per-patient skin-tone measurement. This work demonstrates how the theoretical pulse oximetry model can be applied in calibration.

physics.med-ph↗