Search arXivSearch

arXiv · 2506.21029

Are Ultrathin Stents Optimal for Bifurcation Lesions? Insights from Computational Modelling of Provisional and DK-Crush Techniques

Abstract

Complex coronary bifurcation lesions remain challenging in percutaneous coronary intervention, with stent design and deployment strategy influencing clinical outcomes. This study compares the mechanical and hemodynamic performance of the ultrathin-strut Orsiro and thin-strut Xience Sierra stent in Provisional Side Branch (PSB) and Double Kissing Crush (DKC) techniques. We used finite element analyses of bifurcation stent deployment to assess malapposition, ostium clearance, and arterial wall stress for both techniques. Computational fluid dynamics simulations quantified the luminal exposure to low Time-Averaged Endothelial Shear Stress (TAESS below 0.4 Pa) and high shear rates (above 1000 1/s). In PSB, Orsiro showed higher malapposition (13.0% vs 9.6%) but improved SB ostium clearance (77% vs 64%) and lower low-TAESS exposure (30.3% vs 33.6%) compared to Xience. Orsiro also produced higher arterial wall stresses, particularly during kissing balloon inflation. In DKC, differences in malapposition and ostium clearance diminished between stents, though Orsiro retained a hemodynamic advantage with lower low-TAESS (28.2% vs 36.3%).Stent design influenced outcomes more strongly in PSB, where anatomical interaction and platform-specific behavior impacted both structural and hemodynamic results. In DKC, procedural complexity minimized those differences, making the stenting technique the primary performance driver. Nonetheless, Orsiro consistently preserved more favorable flow conditions. These findings highlight the need to match device selection with lesion characteristics in PSB, while in DKC, optimizing procedural steps may have a greater impact than the choice of stent platform.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Andrea Colombo, Dario Carbonaro, Mingzi Zhang, Chi Shen, Ramtin Gharleghi, Ankush Kapoor, Claudio Chiastra, Nigel Jepson, Mark Webster, Susann Beier. 2025-11-14. Are Ultrathin Stents Optimal for Bifurcation Lesions? Insights from Computational Modelling of Provisional and DK-Crush Techniques. https://doi.org/10.1002/ccd.70329

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Prediction of biological radiation effects based on ionization clusters (nanodosimetry)

This article reviews approaches that link the formation of ionization clusters in nanometric volumes to radiobiological effectiveness. The corresponding models were developed as the field of nanodosimetry developed. Some address early biological radiation effects, such as DNA damage, while most aim to predict cell survival or inactivation. The models also differ in the nanodosimetric quantities considered, with many based on the probability distribution of ionization cluster formation in a single target. Some models account for the synergistic effects of pairs of ionization clusters formed in different targets. Several models feature macroscopic aggregation frameworks based on particle fluence, which are proposed for use in radiotherapy treatment planning, particularly in ion-beam radiotherapy. The models are presented here using harmonized terminology and notation for nanodosimetric quantities. An extension of the conceptual framework of nanodosimetry is also discussed. This extension transitions from a target-centered description to a track-centered description. It also introduces nanodosimetry-based analogs of dosimetric concepts, such as dose and linear energy transfer. This paper traces and summarizes the historical development of nanodosimetry-based biological effect models and discusses conceptual aspects of the models to reveal their underlying assumptions and the extent to which they are mechanistic or merely elucidate correlations. Eventually, an attempt is made to identify the key open questions in this field that still need to be addressed.

physics.med-ph

Contextual Cellular Growth (ConCeG) of neural cells for realistic grey matter tissue generation for diffusion MRI simulations

Accurate interpretation of diffusion magnetic resonance imaging (dMRI) signals in grey matter (GM) remains challenging due to the complex, heterogeneous, and densely packed cellular environment. Numerical phantoms provide a controlled framework for investigating the relationship between microstructure and diffusion signals, yet existing approaches often lack the morphological realism and multi-cellular organisation required to faithfully represent GM tissue. In this work, we introduce Contextual Cellular Growth (ConCeG), a generative framework for creating individual cells or constructing dense, three-dimensional, multi-cellular GM substrates informed by real neuronal and glial morphologies. The method combines topological neuron synthesis with a spatially constrained growth network, allowing for the controlled generation of heterogeneous cellular environments with realistic intra- and extracellular compartments. Synthetic cells are generated using morphological and topological characteristics derived from biological reconstructions. We validate the framework through comparisons of structural features with real cellular data, demonstrating strong agreement in branch order, length, angle, and tortuosity distributions. Power spectrum analysis further shows that both intracellular compartments reproduce the spatial correlations observed in biological tissue. Together, these results show ConCeG provides a biologically grounded framework for generating grey matter substrates suitable for large scale diffusion MRI simulation.

physics.med-ph

Magnetic Field of Firing Neuron in Humans: Measurable by Quantum Sensing MRI?

Firing neurons generate action potentials that propagate along axons to transmit signals supporting cognitive functions. These electrical currents generate magnetic fields, yet direct detection of these neuronal magnetic fields by MRI remains elusive. This Mini Review investigates why this goal has been proven difficult to achieve and whether an emerging approach, quantum sensing MRI, can overcome the challenge.

physics.med-ph