Search arXivSearch

arXiv · 2507.16282

The Automatic Calibration Method of the Compton Edge Based on Normalized Cross-correlation and Simulated Annealing Algorithm

Abstract

Accurate energy channel calibration in scintillation detectors is essential for reliable radiation detection across nuclear physics, medical imaging, and environmental monitoring. Organic scintillators like BC408 and EJ309 lack full-energy peaks, making their Compton edge a critical calibration alternative where traditional peak methods fail. Existing Compton edge identification techniques - Gaussian fitting for the 50%-70% amplitude point, first derivative minimum detection, and Monte Carlo simulation - suffer significant degradation from low count rates, spectral overlap, and subjective interval selection. For the first time, we propose an automated calibration procedure based on Normalized Cross-Correlation (NCC), Simulated Annealing (SA), and a convolutional response model to address these issues. This method automates the selection of the Compton edge interval through NCC-based matching, utilizes SA for global parameter optimization, and then employs a convolutional model for precise matching. Experiments involving the irradiation of organic scintillators (BC408, EJ309) and inorganic scintillators (NaI:Tl, LaBr3:Ce) with 137Cs, 22Na, 54Mn, and 60Co radiation sources demonstrate that this method achieves accuracy commensurate with full-energy peak calibration method (cosine similarity >99.999%) and exhibits superior stability compared to the two traditional methods. In the extreme cases of spectral overlap and low count rate, the average errors of this method are 19.77% and 15.65% of those from the two traditional methods in BC408, 56.44% and 33.15% of those from the two traditional methods in EJ309. This work advances detector calibration and offers a scalable, automated solution for high-energy experiments and portable devices.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Dehua Kong, Yanbiao Zhang, Zixi Lin, Yehao Qiu, Xiulian Chen, Zhonghai Wang. 2025-07-22. The Automatic Calibration Method of the Compton Edge Based on Normalized Cross-correlation and Simulated Annealing Algorithm. https://arxiv.org/abs/2507.16282

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

The Remote Analog to Digital Conversion DAQ System for the TRISTAN Detector Upgrade

The TRISTAN detector is an upgrade to the KATRIN experiment to enable a differential measurement of the tritium $β$-decay spectrum to search for sterile neutrinos with keV masses. This entails performing precision electron spectroscopy with over one thousand silicon drift detector pixels, each responsible for recording incident electron rates of $10^5$ counts per second. A project specific data acquisition (DAQ) system is developed to meet the experimental challenges through a remote analog to digital conversion (RADC) design. In this work, the conceptual design of the RADC DAQ is presented along with the built system for operating the TRISTAN detector upgrade. The system includes flexible signal processing logic and data management that is optimized for the high-rate precision measurement.

physics.ins-det

True Alternating Current Scanning Tunneling Microscope (ACSTM): tunneling on insulators

Scanning Tunneling Microscopy (STM) has revolutionized our atomic scale understanding of surfaces and accelerated progress in nanotechnology. This technique, however, is restricted to metal or semiconducting samples, as it requires a tiny current to stabilize the tip-sample distance with atomic scale precision. We developed a new imaging and feedback method that relies on true alternating current (AC) without any direct current (DC) component. This technique does not only enable the imaging on non-conducting surfaces with atomic step resolution, like (thin) glass and oxides, it provides also access to high-frequency electronic signal coming from the sample. We demonstrate that it is possible to measure on 25nm thick silicon oxide with 10 MHz tunneling current.

physics.ins-det

Charged-particle topology reconstruction with an in-liquid SiPM array

Liquid scintillator detectors instrumented with photosensors inside the scintillation volume preserve local optical information that is largely lost in conventional boundary-readout geometries. We demonstrate that this information is sufficient for charged-particle topology reconstruction using a sparse three-dimensional lattice of silicon photomultipliers. After validating the Geant4 detector response against measured photon-count distributions, a simulation-trained, time-informed convolutional neural network reconstructs the entry and exit points of through-going muons with median residuals of 1.91~cm and 2.39~cm, respectively. The reconstructed endpoints are geometrically consistent with acceptance regions defined by external trigger counters in cosmic-ray muon data. The same framework also reconstructs the production vertices of simulated positron starting-track events with a median residual of about 4.5~cm. These results establish the feasibility of topology-sensitive reconstruction using sparse in-liquid photosensor arrays in homogeneous liquid scintillator detectors.

physics.ins-det