High-Resolution Directional Depth Electrodes: Open-Source FEM Lead-Field Modeling, Characterization, and Validation

Abstract

Depth electrodes used in stereoelectroencephalography (sEEG) and deep-brain stimulation (DBS) are essential tools for neural recording and stimulation. Traditional designs have limited spatial resolution, typically 8 to 16 cylindrical contacts (0.8 to 1.0 mm diameter) along a 5 to 10 cm shaft, restricting recordings from small or localized populations. Recent high-density, directional electrodes (HDsEEG) enable finer localization of local field potentials (LFPs) and spike timing. Yet, characterizing their directional sensitivity and validating modeling tools for lead field (LF) analysis remain critical. We compare finite element method (FEM) LF modeling of a novel HD-sEEG electrode using two tools: a commercial solver (ANSYS) and an open-source pipeline (Brainstorm-DUNEuro). Goals: (i) validate against analytical solutions, (ii) assess solver differences, and (iii) characterize HDsEEG directional sensitivity. LFs were modeled in simple and bio-relevant scenarios. Using Helmholtz reciprocity, we computed LFs by (a) electrode-based stimulation with ANSYS and (b) source-based recording with Brainstorm-DUNEuro. First, a multi-sphere head model with known solutions tested the solver's accuracy. Next, an HDsEEG electrode in a homogeneous conductor was simulated. Directional effects were assessed by comparing sensitivity with vs. without the insulating substrate. Source localization performance was also compared between HDsEEG and standard electrodes. Both solvers closely matched analytic solutions. In realistic settings, LF distributions were highly similar. Modeling showed clear directional sensitivity: contacts facing a source had higher sensitivity than those shadowed, reflecting a substrate shielding effect that vanished when the substrate was conductive. Crucially, HDsEEG improved source localization, as voltage differences across contacts provided robust directional LF information.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Takfarinas Medani, Jace Willis, Chris Wright, Yash Vakilna, Ryan Shores, Raymundo Cassani, Anand Joshi, Richard Leahy, John Seymour, John Mosher. 2025-08-16. High-Resolution Directional Depth Electrodes: Open-Source FEM Lead-Field Modeling, Characterization, and Validation. https://arxiv.org/abs/2508.13212

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Explicit pixel-value correction of lateral response artifacts in radiochromic film dosimetry: Evaluation of single-, double-, and triple-channel methods

Background:Flatbed scanners introduce a lateral response artifact (LRA) perpendicular to lamp motion. Because LRA varies with position, pixel value (PV), color channel, and film model, it may violate the channel-common perturbation assumption of multichannel dosimetry. Purpose:To characterize PV-domain LRA in EBT4 and EBT-XD films, compare Lewis and Full corrections, and evaluate their effects on single-, double-, and triple-channel dosimetry. Methods:Seven films were scanned at seven lateral positions using a cyclic-shift design. For each position and RGB channel, center-equivalent response was modeled as PVc = A + B * PV. Lewis correction used two films with linear interpolation of A and B; Full correction used all center-local pairs with PCHIP interpolation of PVc - PV. Performance was assessed by portrait-landscape profile consistency and dose accuracy. Results: LRA was affine in PV, asymmetric about scanner center, and dependent on film model and channel. Both corrections reduced profile differences. Mean absolute profile differences across MU levels ranged from 1.4%-5.3% for EBT4 and 2.4%-6.8% for EBT-XD. The methods performed similarly for EBT4, while Full correction reduced discrepancies for several EBT-XD double- and triple-channel methods. Residual blue-channel errors limited blue, red-blue, and triple-channel performance. At 500 MU and above, red showed the best dose agreement; at 100 MU, both corrections increased deviations for several single-channel estimates. Full correction produced smaller deviations than Lewis in several EBT-XD conditions. Conclusions:LRA should be corrected in the raw-PV domain before dose reconstruction. Full correction better preserved profile consistency and dose accuracy, particularly for EBT-XD. Because improved profile consistency did not always improve central-dose agreement, both should be evaluated separately in commissioning.

physics.med-ph↗

Integration of Spectral CT with PET and SPECT: Bringing Tissue Composition Information to Molecular Imaging

Molecular imaging has been transformed by integrating positron emission tomography (PET) or single-photon emission computed tomography (SPECT) with x-ray computed tomography (CT). However, x-ray CT in hybrid imaging is used primarily for anatomical localization and for attenuation and scatter correction of emission data. Its ability to characterize tissue composition remains underutilized. Spectral CT, including dual-energy CT (DECT) and photon-counting CT (PCCT), can provide material-specific information, such as images of iodine concentration, bone or calcium fractions, electron-density, effective atomic number and other spatially-varying material properties. These capabilities create an opportunity to bring tissue-composition information into molecular imaging. In addition, spectral CT can also reduce CT artifacts and provide more accurate corrections for emission data. In this review, we discuss the technical basis, integration pathways, and translational opportunities for combining spectral CT with molecular imaging. We review sequential and integrated PET/DECT approaches, emerging PET/PCCT concepts, PET-enabled spectral CT, and extensions to SPECT/CT. We then examine how spectral CT information may improve attenuation correction, scatter correction, tissue-fraction correction, positron range correction, reconstruction priors, and dosimetry. Clinical and translational applications include contrast-enhanced molecular imaging, vascular and perfusion imaging, quantitative bone marrow imaging, musculoskeletal imaging, and theranostics. Overall, spectral CT integration may move hybrid molecular imaging toward tissue-composition-informed molecular imaging, in which radiotracer signals are interpreted within their material context.

physics.med-ph↗

Non-circular scan trajectories for reducing cone-beam artifacts in Gamma Knife CBCT images: a simulation study

Objective. Gamma Knife cone-beam computed tomography (CBCT) images are deteriorated by cone-beam artifacts whose magnitude increases along the superior direction. In this study, a novel scan trajectory compatible with the Gamma Knife CBCT system is optimized to reduce cone-beam artifacts, with the aim to replace the current 200-degree single-arc scan. Approach. Data sampling analysis with tomographic incompleteness maps indicates the level of undersampling across the field of view based on the geometry of the system and of a scan trajectory. Moreover, CBCT simulations are performed from a virtual phantom with disks aligned along the axial direction and from a CT reconstruction of a stereotactic end-to-end validation (STEEV) phantom. CBCT projections are simulated for a given scan trajectory through a polychromatic forward projection model with added noise and scatter, then the CBCT image is reconstructed using an iterative algorithm which minimizes weighted least squares. Main results. Both the incompleteness analysis and the CBCT simulations indicate adding lines to the current single-arc trajectory is more efficient to reduce cone-beam artifacts than adding more arcs, both in terms of number of additional projections and new artifacts. A line-arc-line trajectory with source axial steps of 3.5 mm removes virtually all cone-beam artifacts. The widths of the cone-beam artifacts created by the disks show a positive correlation between the artifact magnitude and the incompleteness value. Significance. A line-arc-line scan trajectory is promising to reduce cone-beam artifacts of the Gamma Knife CBCT images while being a compatible and reasonable trajectory for the current system design.

physics.med-ph↗