A pilot cohort study of a microfluidic-based point-of-care bilirubin measurement system

Abstract

Objective The concentration of bilirubin in blood or serum is useful for assessing liver function as well as monitoring treatment. This study evaluates the clinical performance of a novel point-of-care (PoC) device for the detection of bilirubin in serum. The PoC device incorporates an integrated miniature optoelectronic sensing module and a microfluidic test cartridge. Methods Patients' serum total bilirubin concentrations, ranging from 2 μmol/L to 480 μmol/L, were measured using the PoC device and the standard laboratory method (n=20). Bland-Altman analysis and regression analysis using Passing-Bablok method were used to benchmark the PoC device against the standard laboratory measurements. The diagnostic capability of the PoC device in categorising the serum samples within clinically relevant bilirubin concentration thresholds of 200, 300, and 450 μmol/L was assessed using receiver operating characteristic (ROC) analysis. Results The mean difference between the PoC device and the standard laboratory method was -5.6 μmol/L, with a 95% confidence interval (CI) of -45.1 μmol/L to 33.9 μmol/L. The coefficient of determination (R2) was 0.986. The PoC device achieved a detection sensitivity of 90% and specificity of 97% in categorising bilirubin concentrations within bands used in clinical decision-making. Conclusions This study demonstrates that the proposed PoC device is capable of measuring bilirubin levels in patient samples with clinically acceptable accuracy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jean Pierre Ndabakuranye, Inge W. G. Last, Kay Weng Choy, Peter Thurgood, Jason C. Steel, Genia Burchall, Stella Stylianou, Khashayar Khoshmanesh, Arman Ahnood. 2025-10-10. A pilot cohort study of a microfluidic-based point-of-care bilirubin measurement system. https://doi.org/10.1016/j.lmd.2025.100073

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Prediction of biological radiation effects based on ionization clusters (nanodosimetry)

This article reviews approaches that link the formation of ionization clusters in nanometric volumes to radiobiological effectiveness. The corresponding models were developed as the field of nanodosimetry developed. Some address early biological radiation effects, such as DNA damage, while most aim to predict cell survival or inactivation. The models also differ in the nanodosimetric quantities considered, with many based on the probability distribution of ionization cluster formation in a single target. Some models account for the synergistic effects of pairs of ionization clusters formed in different targets. Several models feature macroscopic aggregation frameworks based on particle fluence, which are proposed for use in radiotherapy treatment planning, particularly in ion-beam radiotherapy. The models are presented here using harmonized terminology and notation for nanodosimetric quantities. An extension of the conceptual framework of nanodosimetry is also discussed. This extension transitions from a target-centered description to a track-centered description. It also introduces nanodosimetry-based analogs of dosimetric concepts, such as dose and linear energy transfer. This paper traces and summarizes the historical development of nanodosimetry-based biological effect models and discusses conceptual aspects of the models to reveal their underlying assumptions and the extent to which they are mechanistic or merely elucidate correlations. Eventually, an attempt is made to identify the key open questions in this field that still need to be addressed.

physics.med-ph

Contextual Cellular Growth (ConCeG) of neural cells for realistic grey matter tissue generation for diffusion MRI simulations

Accurate interpretation of diffusion magnetic resonance imaging (dMRI) signals in grey matter (GM) remains challenging due to the complex, heterogeneous, and densely packed cellular environment. Numerical phantoms provide a controlled framework for investigating the relationship between microstructure and diffusion signals, yet existing approaches often lack the morphological realism and multi-cellular organisation required to faithfully represent GM tissue. In this work, we introduce Contextual Cellular Growth (ConCeG), a generative framework for creating individual cells or constructing dense, three-dimensional, multi-cellular GM substrates informed by real neuronal and glial morphologies. The method combines topological neuron synthesis with a spatially constrained growth network, allowing for the controlled generation of heterogeneous cellular environments with realistic intra- and extracellular compartments. Synthetic cells are generated using morphological and topological characteristics derived from biological reconstructions. We validate the framework through comparisons of structural features with real cellular data, demonstrating strong agreement in branch order, length, angle, and tortuosity distributions. Power spectrum analysis further shows that both intracellular compartments reproduce the spatial correlations observed in biological tissue. Together, these results show ConCeG provides a biologically grounded framework for generating grey matter substrates suitable for large scale diffusion MRI simulation.

physics.med-ph

A wearable, stretchable radio-frequency coil for extremity imaging in low-field MRI systems

Purpose: To develop a wearable, stretchable radio-frequency (RF) coil that conforms to the extremities and improves the filling factor in low-field MRI. Methods: A stretchable solenoid RF coil was fabricated by stitching a sinusoidally arranged Litz-wire conductor onto an elastic textile. The coil was compared with four rigid coils, including two routinely used designs and two prototypes designed to isolate the effects of conductor material and stretchability. Performance was characterized through quality-factor, loading-factor, transmit-efficiency, and image-based signal-to-noise ratio (SNR) measurements. Phantom and in-vivo knee experiments for one volunteer were performed using two portable MRI systems operating at 3.53 and 3.04 MHz. Results: The electrical performance of the selected Litz wire decreased with increasing frequency, becoming a slight disadvantage at 3.53 MHz. Nevertheless, the stretchable coil provided the highest phantom SNR in both systems, exceeding that of the best-performing rigid coil by approximately 8 % at 3.53 MHz and 20 % at 3.04 MHz. In vivo, its global SNR was 5 % lower than that of the best rigid coil at 3.53 MHz and 19 % higher at 3.04 MHz. Compared with the larger rigid coil required when knee positioning is constrained, the corresponding SNR improvements were approximately 18 % and 80 %. Conclusion: Anatomical conformity can compensate for the frequency-dependent electrical limitations of Litz wire in stretchable low-field RF coils. The proposed design provided SNR comparable to or greater than the best rigid designs while facilitating coil placement in subjects for whom smaller rigid coils may be impractical.

physics.med-ph