Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis

Abstract

A systematic study is conducted to understand the coincident resolving time (CRT) for a pair of Lutetium-yttrium oxyorthosilicate (LYSO) and the plastic scintillation detector bars under the Geant4 framework. Crystals are coupled to a silicon photomultiplier single pad wafer with an appropriate optical coupling for signal readout. Pad reads the light photons undergoing optoelectronic conversion at the wafer site and generates electrical pulses with a bi-exponential shape. These signals are used to determine the trigger time stamp of back-to-back gamma-rays emitted from a point source, enabling the evaluation of CRT performance at different plastic scintillator lengths. For the LYSO detector, the simulation yields the CRT response of 174 ps, which aligns to the literature reported value for the dimensions 2 mm x 2 mm x 10 mm. To identify the plastic scintillator dimensions with an integrated gamma-ray detection efficiency comparable to LYSO's photopeak efficiency at 511 keV gamma photons, various bar lengths of commercial plastic BC404 and TIFR Ooty's in-house developed plastic material are attempted in Geant4. Consequently, for both the plastic scintillators, the equivalent length (for the same cross-sectional area) was found to be 4 times that of LYSO crystal length at a threshold of 25 keV. CRT value determined for this dimension is found to be $\approx$ 300 ps for both the plastic medium. It suggests that, if an animal preclinical PET scanner is developed with plastic bars, the minimum achievable image resolution (FWHM) of $\approx$ 4.5 cm can be expected for a pair of detection elements.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Davinder Siwal, P. K. Mohanty, D. Bose, A. Jain. 2025-10-16. Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis. https://arxiv.org/abs/2510.14437

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Mortality of ultra-thin LGADs and PiN diodes from high energy deposition

Low Gain Avalanche Diodes are prime candidates for high-resolution timing applications in High Energy Physics, Nuclear science, and several other fields. Operating these devices in high-radiation environments presents various hazards, including the risk of their permanent degradation or destruction caused by effects such as Single Event Burnout. Studies using minimum ionizing particles found a greatly reduced Single Event Burnout risk by operating below a bias voltage corresponding to an average electric field of 12 V/$μ$m - however, as high energy particle colliders produce a wide energy spectrum of radiation, it is crucial to understand this phenomenon and other possible damage mechanisms at energy deposition levels greater than those of minimum ionizing particles. This was achieved by pre-irradiating LGADs and PiN diodes with active thicknesses of 20, 30, and 50 $μ$m up to 1.5 $\times$ 10$^{15}$ $\mathrm{n_{eq}/cm^2}$, and exposing them to beams of protons and heavy ions (C, O, Fe, Au) at the BNL Tandem van de Graaff accelerator. Several mortality categories were observed, defined by different electrical and mechanical damage signatures. This furthers our understanding of permanent radiation damage of silicon devices, crucial towards mitigating Single Event Burnout and other damage mechanisms to safely operate future detectors.

physics.ins-det

Small but Tubby: A Magnetic Loop Antenna Made from 100 mm Copper Tubing

This paper presents the electrical model, key equations, and practical construction of a small transmitting magnetic loop antenna built from unusually large 100 mm diameter copper tubing. The large conductor surface area and wide-area transitions to the vacuum capacitors were designed to minimize resistive losses. The frequency range from 1.8 MHz to 31 MHz is unusually wide. Frequency, impedance matching, and azimuth are all adjusted automatically by servo motors. Indoor losses originate predominantly from near-field coupling to the environment rather than from the antenna itself. Temperature-rise measurements confirm that the bulk of the dissipated power is absorbed by the environment, not by the antenna components. The conducted H-field measurements demonstrate good agreement between the measured H-field and the theoretical free-space H-field calculated from the antenna geometry and an estimated loop current. The loop current was estimated from the measured antenna bandwidth and the applied transmit power. The antenna was developed for indoor operation where outdoor installation is not possible.

physics.ins-det

Construction and characterisation of the DarkSide-20k veto silicon photo-multiplier tiles

Silicon photo-multipliers (SiPMs) are state-of-the-art sensors capable of detecting a single photoelectron under cryogenic conditions, with potentially lower radioactivity than widely used photomultiplier tubes. The DarkSide-20k experiment, designed to perform direct dark matter searches using liquid argon as the target material, employs SiPM technology to detect interactions in the active detector volumes, including the central dual-phase Time Projection Chamber and the Inner and Outer Veto volumes. The vetoes are designed to discriminate against radiogenic neutron and cosmic muon backgrounds associated with the dark matter search. This paper describes the completed production and test protocols for the "Veto Tiles" (called vTiles, arrays of 24 SiPMs integrated on a printed circuit board providing the power distribution and signal amplification); 16 vTiles are grouped into "Veto Photo-Detector Units" to instrument the Inner Veto volume. Each vTile underwent detailed testing at room and cryogenic temperatures, confirming stable operation, high signal-to-noise ratio, and low radioactive contamination, demonstrating the robustness of the proposed design for cryogenic conditions. The final production yield exceeded 87%, surpassing the 80% requirement and corresponding to 1920 Veto Tiles to populate 120 Veto Photo-Detector Units, plus an additional 6% as spares.

physics.ins-det