Abstract
Non-destructive beam diagnostics are essential for low-energy medical cyclotrons, where even thin interceptive devices can severely degrade beam quality. We investigate an external fiber monitor (EFM) based on Ce-doped silica scintillating fibers that detects secondary radiation generated at existing beamline components of the 18 MeV Bern Medical Cyclotron beam transfer line (BTL). Three use cases were studied: (i) beam intensity monitoring around an electrically isolated, water-cooled beam dump; (ii) beam-loss monitoring around a 10 mm collimator under varying the beam focusing; and (iii) by steering a 6.5 mm $\times$ 6.5 mm beam spot on a beam dump. For case (i), the summed EFM signal exhibits a linear dependence on the current on target over nearly three orders of magnitude. In case (ii), a normalized EFM-based beam-loss proxy scales monotonically with an electrical loss proxy across several focusing settings. Furthermore, opposing-fiber signal ratios provide decoupled, monotonic sensitivity to horizontal and vertical beam displacements.
Explore related subjects
Keep this discovery
Alexander Gottstein, Pierluigi Casolaro, Gaia Dellepiane, Lars Eggimann, Eva Kasanda, Isidre Mateu, Samuel Usherovich, Paola Scampoli, Cornelia Hoehr, Saverio Braccini. 2026-01-26. Non-Destructive Beam Monitoring via Secondary Radiation Detection with Ce-Doped Silica Fibers. https://doi.org/10.1088/1748-0221/21/03/p03036
Cite the original work for its findings. Save a collection to share your selection of sources.