Liver Metastasis Detection at Reduced Radiation Dose: Diagnostic Evaluation of a Novel Organ-Level Tube Current Modulation Method

Abstract

Objective: To evaluate retroOpt, a novel organ-level tube current modulation (TCM) method that minimizes effective radiation dose while preserving diagnostic image quality for liver metastasis detection. Methods: In this retrospective, IRB-approved study, 22 patients with 68 liver lesions (38 malignant, 30 benign) underwent portal venous phase contrast-enhanced chest-abdomen-pelvis CT. Using projection-domain noise emulation, five series were generated per patient: original full dose (Orig-100), uniform dose reduction to 40% (UD-40) and 60% (UD-60) of the original effective dose, and organ-level TCM at the same levels (Opt-40, Opt-60). Three abdominal radiologists independently detected and classified lesions and rated image quality (5-point Likert). Per-lesion sensitivity was compared by McNemar test. Malignant lesion-size thresholds for 50% and 90% sensitivity (x50, x90) were estimated by logistic regression. Image quality was compared by Wilcoxon signed-rank test. Results: At the 60% level, Opt-60 achieved mean malignant lesion sensitivity comparable to full-dose CT (82% vs 80%) and exceeded uniform reduction (67%). At the 40% level, Opt-40 improved mean sensitivity over UD-40 from 52% to 68%. Mean all-lesion sensitivity rose from 54% (UD-40) to 64% (Opt-40) and from 65% (UD-60) to 77% (Opt-60), remaining comparable to full dose (76%). All optimized-versus-uniform differences were significant (p <= 0.003). Opt-60 lesion-size thresholds closely matched Orig-100 (x50 3.3 vs 3.7 mm; x90 15.7 vs 16.8 mm). Median Opt-60 image quality ranged from 3 to 4.5 across readers. Conclusions: Organ-level TCM preserved liver metastasis detection at 60% of the original effective dose, outperforming uniform dose reduction. Task-specific organ-level dose optimization may enable greater CT dose reduction than uniform strategies for patients requiring repeated metastasis surveillance.

Explore related subjects

Keep this discovery

BibTeXRIS

Maria Jose Medrano, Sen Wang, Kaylee W. Fang, Jennie Cao, Liyan Sun, Luyao Shen, Andy Shon, Luke Yoon, Grant M. Stevens, Adam S. Wang, Justin R. Tse. 2026-07-28. Liver Metastasis Detection at Reduced Radiation Dose: Diagnostic Evaluation of a Novel Organ-Level Tube Current Modulation Method. https://arxiv.org/abs/2607.25440

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph