Relative and absolute dosimetric commissioning of the ELIMAIA--ELIMED laser-driven proton beamline at 23.4 MeV

Abstract

The development of laser-driven proton beamlines for biomedical and radiobiological applications requires traceable dosimetry and reliable online monitoring at the irradiation point. In this work, we report the relative and absolute dosimetric commissioning of the ELIMAIA--ELIMED laser-driven proton beamline at ELI Beamlines using an energy-selected proton beam with an average energy of about 24~MeV. Radiochromic-film measurements were used to characterize the transverse dose distribution, the depth--dose profile, and the proton energy spectrum at the irradiation point. The reconstructed spectrum was centred at $23.45~\mathrm{MeV}$ with a FWHM of $2.60~\mathrm{MeV}$, while the transverse dose distribution showed an approximately $5.5~\mathrm{mm}$ field size with a millimetre-scale homogeneous region. The Faraday Cup was used as the absolute reference detector for dose to water determination and for cross-calibrating the Dual-Gap Ionization Chamber, which was operated as the primary online dose monitor. The Integrating Current Transformer and Secondary Electron Monitor were evaluated as upstream relative beam monitors. An independent RCF--FC dose comparison performed over fifty consecutive shots yielded $30.45 \pm 3.5~\mathrm{cGy}$ from the FC and $36.46 \pm 1.8~\mathrm{cGy}$ from the EBT3 film. Dedicated G4ELIMED Monte Carlo simulations showed that the RCF positioned upstream of the FC perturbs the FC measurement through proton losses outside the FC acceptance and through an increase in the effective beam area at the FC entrance. Applying the resulting correction factor, $C_{\mathrm{MC}}=1.109$, reduced the residual difference between the FC- and RCF-derived doses to about $7\%$. These results establish the dosimetric chain of the ELIMAIA--ELIMED beamline under the present low-fluence commissioning conditions and identify the operational limits of the online monitoring system.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

G. A. P. Cirrone, G. Petringa, A. Kurmanova, R. Catalano, A. Amato, A. Pappalardo, M. G. Guarrera, F. Schillaci, P. Blaha, F. Grepl, M. Tryus, A. Velyhan, L. Giuffrida, D. Margarone, G. Cuttone. 2026-09-09. Relative and absolute dosimetric commissioning of the ELIMAIA--ELIMED laser-driven proton beamline at 23.4 MeV. https://arxiv.org/abs/2608.15272

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-power attosecond X-ray free-electron lasers: physics and design strategy

Attosecond pulses from X-ray free-electron laser (XFEL) have opened new opportunities for probing ultrafast electronic dynamics on the Angstrom--attosecond spatiotemporal scale. Most attosecond XFEL concepts rely on generating an ultrashort high-current spike through either external laser modulation or accelerator-based beam manipulation. Despite their different implementations, these approaches share the same essential physics, namely that the XFEL amplification is confined to a short effective lasing window within the electron beam. However, existing studies are often scheme-specific and do not yet provide a unified quantitative picture of how fundamental electron-beam properties constrain high-power attosecond performance. In this work, we investigate the general physics and scheme-independent requirements for generating high-power attosecond X-ray pulses from a short current spike. From the perspective of post-saturation superradiant evolution, we show that the effective lasing length of the electron beam governs both the attainable peak power and the pulse duration. We further examine the distinct roles of slice energy spread, slice emittance, energy chirp, undulator tapering, and transverse beam tilt. Our results reveal the trade-off between peak power, pulse shortening, and single-spike probability, and provide facility-independent guidelines for optimizing electron-beam phase-space manipulation toward terawatt-class attosecond XFEL operation.

physics.acc-ph

Bayesian Optimization of The Relativistic Heavy Ion Collider Luminosity via $s^*$ Control

Maximizing luminosity at the interaction point (IP) requires the collision location $s_{IP}$ to coincide with the longitudinal position of the minimum beta function, $s^$. Accurate optics measurements and control of $s^$ are therefore essential for luminosity optimization. At the Relativistic Heavy Ion Collider (RHIC), average horizontal beta-beat measurements between operating IPs are approximately $20%$, with significant variation in measured $s^*$. Precise control of the beam waist position $s^$ is particularly challenging for modern high-energy colliders with short bunch lengths and large crossing angles. We present an online Bayesian optimization (BO) application using the GPTune framework to optimize sPHENIX luminosity through $s^$ control at RHIC. GPTune was first validated at the RHIC Electron Beam Ion Source (EBIS), where it achieved up to a $70%$ increase in beam intensity over the baseline, although experienced operators could reach similar performance with longer manual tuning. The framework was subsequently deployed during sPHENIX operations. Using an intensity-normalized Zero-Degree Calorimeter (ZDC) signal as the optimization objective, due to the unavailability of the live sPHENIX MVTX signal, GPTune identified local luminosity maxima, recovered from intentionally degraded $s^$ configurations, and revealed residual horizontal and vertical waist offsets in the interaction region. These results demonstrate the robustness and efficiency of Bayesian optimization for real-time collider tuning under noisy, time-varying conditions. The $s^$ control methodology provides a promising tool for precision luminosity optimization and is particularly relevant to next-generation short-bunch colliders such as the Electron-Ion Collider (EIC).

physics.acc-ph

Ultralow Mean Transverse Energy and High Quantum Efficiency Cryogenic Bialkali Photocathode for MHz-Repetition-Rate Electron Sources

Simultaneously achieving high quantum efficiency (QE), ultralow mean transverse energy (MTE), and robust long-term operation under conditions relevant to continuous-wave (CW) X-ray free-electron lasers (XFELs) remains a central challenge for semiconductor photocathodes. This challenge arises from the trade-off between QE and MTE, as well as the difficulty of maintaining stable operation in high-field CW electron guns. Here we demonstrate a cryogenic K2CsSb photocathode that simultaneously achieves high QE, ultralow MTE, and robust long-term operation in a CW gun under XFEL-relevant operating conditions. Under cryogenic operation, the photocathode achieves an MTE of 50 meV while sustaining a QE of 5.4%. Milliampere-level CW current, including operation at 5 mA, was demonstrated together with an approximately 20-day operational history. The observations are consistent with improved carrier survival and/or surface escape in photocathodes prepared using the optimized recipe. These results show that the practical QE-MTE trade-off can be substantially mitigated in cryogenic bialkali photocathodes and provide a practical pathway toward high-brightness electron sources for CW XFELs and energy-recovery linacs.

physics.acc-ph