Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT

Abstract

Objective. $^{43}$Sc and $^{44}$Sc are positron-emitting isotopes of scandium that have attracted growing interest for theranostic applications in conjunction with the therapeutic isotope $^{47}$Sc. Both isotopes are non-pure positron emitters, and the associated prompt-gamma emission together with the high positron energy introduce complexities in achieving precise quantitative imaging. Approach. We evaluated the quantitative imaging performance of $^{44}$Sc and a $^{43/44}$Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom. We assessed contrast recovery coefficients, image noise and bias for different numbers of iterations and for absolute versus relative scatter correction, and implemented an isotope-specific point-spread-function (PSF) correction based on FLUKA-simulated positron ranges. Main results. The quantitative imaging performance of both isotopes is comparable to $^{18}$F, and the vendor's prompt-gamma correction effectively suppresses the additional random coincidences and lung error caused by prompt-gamma emission. The isotope-specific PSF correction further improves contrast recovery. While the bias obtained for $^{44}$Sc meets the EARL acceptance criterion for non-standard radionuclides, the bias for the $^{43/44}$Sc mixture exceeds it, which we attribute to the mixture-specific cross-calibration of the dose calibrators rather than to the PET imaging. Significance. Our results support the use of LAFOV PET/CT for quantitative imaging of scandium isotopes and provide practical reconstruction recommendations for clinical protocols.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lorenzo Mercolli, Mohammadreza Teimoorisichani, Alenka E. Gaschen, Pascal V. Grundler, Anzhelika N. Moiseeva, Saverio Braccini, Nicholas P. van der Meulen. 2026-10-20. Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT. https://arxiv.org/abs/2610.28884

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Toward opportunistic coronary artery calcium assessment from dual-layer dual-energy chest radiography using deep learning: a simulation study

Background: Coronary artery calcium (CAC) predicts cardiovascular risk. Dual-energy (DE) chest radiography may enable accessible opportunistic CAC assessment by separating calcifications from overlapping bone. Purpose: To evaluate multi-task deep learning for CAC localization and CT-referenced score prediction and assess projection, image representation, and exposure effects. Methods: This retrospective simulation study used 550 calcium-scoring CT exams with reference CAC masks and Agatston scores. Dual-layer DE radiographs were simulated with a physics-based detector model. A modified nnU-Net jointly predicted CAC localization and scores from continuous probability maps. Comparisons included AP, 45°, and lateral views; lateral single-energy (SE), DE bone, and high-pass filtered DE bone images; and dose-matched models at 70, 50, and 24 mAs. Patient-level splitting reserved 10% for independent testing; five-fold models provided averaged inference. Metrics included Dice and AUC at CAC thresholds >0, >100, >400, and >1000, with bootstrap uncertainty and paired DeLong tests with Holm-Bonferroni correction. Results: Lateral views achieved Dice 0.34 versus 0.23 (AP) and 0.22 (45°), with AUCs for CAC >100 of 0.86, 0.81, and 0.77, respectively; AUC differences were not significant after correction. Lateral DE bone outperformed SE (Dice: 0.33 vs. 0.19; AUC for CAC >100: 0.86 vs. 0.64), with significant AUC advantages at >100, >400, and >1000. High-pass filtering added little classification benefit. Dice decreased with exposure, while AUC differences across doses were not significant. Conclusions: Simulation results support lateral dual-layer DE bone imaging for opportunistic CAC assessment, with significant advantages over SE. Physical and clinical validation is needed.

physics.med-ph↗

Three-dimensional stochastic retinal models for synthetic eye-models

Purpose: Current eye models often assume a fixed spherical retinal curvature of 12 mm, which does not reflect the anatomical variability observed in vivo. This study proposes an extension incorporating the stochastic relationship between axial length and three-dimensional (3D) retinal curvature, enabling more realistic 3D eye models. Methods: Axial length and 3D retinal curvature data from 94 subjects, split into five subsets for cross-validation, were used to construct two probabilistic models: a Single Gaussian Model (SGM) and two-component Multi Gaussian Mixture Model (MGMM). Both were converted into conditional forms to estimate retinal curvature for a specific axial length, and evaluated against measured data distributions. Results: All models produced radii of curvature comparable to measured data (median difference <0.2 mm). Non-conditionally, the MGMM described the distribution slightly better than SGM, with higher density overlap (>94% vs. >86%), but similar variance (worst f-test (p-value) 1.05 (0.71) vs. 1.05 (0.73)). Conditioned on axial length, both models performed similarly with overlap exceeding 93% and the worst f-test 1.02 (0.84). Given its simplicity and robustness, the conditional SGM was selected, achieving similar validation results: median difference (<0.10 mm), overlap (>92%), and a non-significant worst f-test (1.02 (0.85)). Conclusions: The conditional SGM enables accurate generation of anatomically realistic 3D retinal curvatures based on axial length, supporting improved stochastic eye modeling for visual optics and population-based simulations of healthy eyes. Translational relevance: Replacing the fixed 12 mm retina assumption with a population-based stochastic model improves anatomical accuracy of eye models used in optical design and treatment planning.

physics.med-ph↗

Sensitivity and Identifiability of Tumor-Related Changes in Longitudinal Microwave Breast Imaging

Microwave imaging could monitor breast tumor response to neoadjuvant chemotherapy, but between-visit changes in breast positioning and normal tissue may confound tumor-related changes. We use eight patient-specific dielectric models from longitudinal MRI of two patients at four time points. For each model, 24-port S-parameters were simulated from 1 to 3 GHz, and Jacobians computed for tumor permittivity and loss, fibroglandular tissue (FGT) permittivity, and breast rotation and translation. 1 mm positioning changes produced changes 20 to 50 times larger than a 10% tumor permittivity change above 2 GHz. Nevertheless, tumor sensitivity patterns were weakly aligned with positioning patterns, tumor-FGT similarity exceeded 0.8 at only some visits, and the Jacobians remained well conditioned. Tumor responses are thus small but separable; monitoring will require positioning correction, frequency selection, and noise-aware detectability analysis.

physics.med-ph↗