Emission-associated energy deposition and Cherenkov-dose proportionality in megavoltage photon beams

Abstract

Background: Quantitative Cherenkov-to-dose conversion depends on beam spectrum, field size, depth, and material properties. Purpose: To distinguish dose contributions with and without accompanying Cherenkov production and assess their implications for calibration. Methods: Geant4 simulations paired dose and local Cherenkov distributions in water for flattened 6, 10, and 15 MV and flattening-filter-free (FFF) 6 and 10 MV beams with $3 \times 3$ and $10 \times 10$ cm$^2$ fields. Energy deposition was tagged by whether its transport step produced at least one Cherenkov photon. Component fractions were calculated over the phantom. Distributions were independently normalized within a shared reference region and compared using voxelwise ratios, lateral profiles, and spatial partitioning. Results: Emission-tagged fractions were 37%-61%, increasing with beam energy and lower for FFF beams. Fractions differed by less than 1.3 percentage points between field sizes. Emission-tagged ratios clustered near unity, with 98.4%-99.9% between 0.95 and 1.05 across the evaluated regions, whereas total and non-emission-tagged ratios varied with depth and lateral position. Emission-tagged profiles generally showed smaller residual differences. Conclusions: Emission-tagged deposition closely tracked local Cherenkov production. Departures of the total Cherenkov-to-dose ratio followed spatial variations in the emission-tagged dose fraction, indicating where calibration needs correction. The classification reflects actual photon production and depends on transport settings; dedicated electron-energy scoring is needed to isolate sub-threshold contributions.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Xianghuan He, Yao Hao, Changyu Sun, Andy N. Wiles, Gregory B. Biedermann, Brian W. Pogue, Rongxiao Zhang. 2026-11-03. Emission-associated energy deposition and Cherenkov-dose proportionality in megavoltage photon beams. https://arxiv.org/abs/2611.05700

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom

Computed tomography (CT) is used ubiquitously for diagnostic imaging, with novel iterative reconstructions such as hybrid iterative reconstruction with partial model-based component (ASiR-V) and deep learning image reconstruction (DLIR) increasingly utilized. In conjunction, CT vendors have developed a noise index (NI) to reduce radiation dose during protocol development. However, it is unclear how these reconstruction algorithms reduce noise as NI changes, particularly across multiple organs. The goal of this study was to utilize an anthropomorphic phantom to evaluate image noise against the prescribed NI across different organs, determine if DLIR reduces image noise compared to ASiR-V, and use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels. Repeat images were acquired on a GE HealthCare Revolution Apex CT scanner across NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku with ASiR-V and DLIR. Organs-of-interest were contoured using TotalSegmentator in 3D Slicer. The correlation between the CTDIvol and NI was determined. The standard deviation of the difference in the repeat images was used as a noise metric (NM), correlated with NI, used to determine the NI at equivalent NM between ASiR-V and DLIR, then applied to determine the dose reduction and difference in NM. The relationship between NI and the resulting CTDIvol was found to be proportional to a power law of exponent -2.23. The maximum potential absolute dose reduction from using DLIR was found in the liver at 1.3 mGy with an NI of 7.9. Across all organs, the relative dose reduction increased, the NM-to-NI ratio decreased, and the noise reduction increased across NI. The relationship between image noise and the prescribed NI differed between ASiR-V and DLIR and varied by organ, with model-based dose reductions of up to 1.3 mGy in the liver.

physics.med-ph↗

Constrained estimation of rotational invariants of the cumulant expansion (RICE) for rapid tensor-valued diffusion MRI

Purpose: To complement 1.5-minute measurements of common tensor-valued diffusion MRI (dMRI) markers with rapid constrained fitting. Methods: Fast dMRI protocols for obtaining rotational invariants of the cumulant expansion (RICE) were paired with constrained weighted linear least squares (CWLLS) to stabilize the more fragile WLLS fit. A compact constraint set was formulated, including a novel mean-dependent upper bound on total diffusional variance. Evaluation used diffusion tensor distribution (DTD) simulations, healthy-volunteer data with a resolution-dependent SNR experiment, and a glioma patient dataset. A 5-minute q-space trajectory imaging (QTI) protocol served as a reference. Results: Across experiments, CWLLS reduced unphysical estimates and fit outliers in parameters such as microscopic FA and isotropic diffusivity variance. In simulations, it narrowed error distributions most clearly in the CSF-dominant case, while some metrics showed a bias-variance trade-off. In vivo, CWLLS removed negative variance estimates, truncated out-of-bounds tails, and reduced artifacts in fluid-contaminated voxels while preserving anatomical contrast. It also retained more stable maps than WLLS at higher resolution, although both estimators degraded in the lowest-SNR setting. Notably, the new mean-dependent variance bound was violated in 15.4% of voxels in the patient dataset, accounting for nearly half of the 32.7% that violated at least one constraint. Healthy-volunteer benchmarking showed that CWLLS completed in under 30 seconds. The constrained QTI fit required 72 minutes, making CWLLS 160 times faster. Conclusion: CWLLS for fast RICE yielded high-quality parameter maps at an online-ready computational cost. This may enhance the reliability of dMRI tissue characterization and strengthen the path toward clinical translation.

physics.med-ph↗

Estimation of Very High Energy Electron (VHEE) beam parameters using an in-air screen

Very high energy electron (VHEE) beams offer several promising properties for radiotherapy. Accurate beam delivery requires a beam model that accounts for scattering along the beam propagation path. This work develops and validates a beam model incorporating Multiple Coulomb Scattering (MCS) to estimate beam parameters. MCS is incorporated as an additive term to the covariance matrix transformation for a drift space, allowing conventional quadrupole scans to estimate beam parameters in the presence of a downstream scatterer. The scattering correction factor is fitted to Monte Carlo (MC) simulations for a range of materials and thicknesses. The model is then validated by recovering the initial beam parameters from simulated quadrupole scans of a realistic beamline. Finally, the method is applied to in-air screen measurements at the Pulsed Energetic Electrons for Research (PEER) end station of the Australian Synchrotron. After fitting the correction factors, the mean and standard deviation of the estimation errors for the final emittance, diameter, divergence, and mean kinetic energy were $(-0.06 \pm 3.17)\%$, $(0.15 \pm 0.66)\%$, $(0.18 \pm 2.18)\%$, and $(-0.03 \pm 0.06)\%$, respectively. Simulated quadrupole scans successfully recovered the initial beam parameters, with errors of at most $13\%$. PEER measurements yielded emittance estimates ranging from $0.238$ to $1.753~\mathrm{mm\,mrad}$ across scans, which were within the expected range. Following an initial quadrupole scan, the method allows the in-air beam size to be estimated for other quadrupole configurations. Further work is required to improve the consistency and reliability of the experimental estimates.

physics.med-ph↗