arXiv · physics/0301008
Investigation of the performance of an optimised MicroCAT, a GEM and their combination by simulations and current measurements
Abstract
A MicroCAT (Micro Compteur A Trous) structure which is used for avalanche charge multiplication in gas filled radiation detectors has been optimised with respect to maximum electron transparency and minimum ion feedback. We report on the charge transfer behaviour and the achievable gas gain of this device. A three-dimensional electron and ion transfer simulation is compared to results derived from electric current measurements. Similarly, we present studies of the charge transfer behaviour of a GEM (Gas Electron Multiplier) by current measurements and simulations. Finally, we investigate the combination of the MicroCAT and the GEM by measurements with respect to the performance at different voltage settings, gas mixtures and gas pressures.
Explore related subjects
Keep this discovery
A. Orthen, H. Wagner, H. J. Besch, R. H. Menk, A. H. Walenta, U. Werthenbach. 2003-01-06. Investigation of the performance of an optimised MicroCAT, a GEM and their combination by simulations and current measurements. https://doi.org/10.1016/s0168-9002(03)00338-3
Cite the original work for its findings. Save a collection to share your selection of sources.