arXiv · 2403.12331
Deep Few-view High-resolution Photon-counting CT at Halved Dose for Extremity Imaging
Abstract
X-ray photon-counting computed tomography (PCCT) for extremity allows multi-energy high-resolution (HR) imaging but its radiation dose can be further improved. Despite the great potential of deep learning techniques, their application in HR volumetric PCCT reconstruction has been challenged by the large memory burden, training data scarcity, and domain gap issues. In this paper, we propose a deep learning-based approach for PCCT image reconstruction at halved dose and doubled speed validated in a New Zealand clinical trial. Specifically, we design a patch-based volumetric refinement network to alleviate the GPU memory limitation, train network with synthetic data, and use model-based iterative refinement to bridge the gap between synthetic and clinical data. Our results in a reader study of 8 patients from the clinical trial demonstrate a great potential to cut the radiation dose to half that of the clinical PCCT standard without compromising image quality and diagnostic value.
Explore related subjects
Keep this discovery
Mengzhou Li, Chuang Niu, Ge Wang, Maya R Amma, Krishna M Chapagain, Stefan Gabrielson, Andrew Li, Kevin Jonker, Niels de Ruiter, Jennifer A Clark, Phil Butler, Anthony Butler, Hengyong Yu. 2024-03-19. Deep Few-view High-resolution Photon-counting CT at Halved Dose for Extremity Imaging. https://doi.org/10.1109/tmi.2025.3618754
Cite the original work for its findings. Save a collection to share your selection of sources.