Calibration of the Cryogenic Measurement System of a Resonant Haloscope Cavity
Abstract
Possible light bosonic dark matter interactions with the Standard Model photon have been searched by microwave resonant cavities. In this paper, we demonstrate the cryogenic readout system calibration of a 7.138 GHz copper cavity with a loaded quality factor $Q_l=10^4$, operated at 22 mK temperature based on a dilution refrigerator. Our readout system consists of High Electron Mobility Transistors as cryogenic amplifiers at 4 K, plus room-temperature amplifiers and a spectrum analyzer for signal power detection. We test the system with a superconducting two-level system as a single-photon source in the microwave frequency regime and report an overall 95.6 dB system gain and -71.4 dB attenuation in the cavity's input channel. The effective noise temperature of the measurement system is 7.5 K.
Explore related subjects
Keep this discovery
Dong He, Jie Fan, Xin Gao, Yu Gao, Nick Houston, Zhongqing Ji, Yirong Jin, Chuang Li, Jinmian Li, Tianjun Li, Shi-hang Liu, Jia-Shu Niu, Zhihui Peng, Liang Sun, Zheng Sun, Jia Wang, Puxian Wei, Lina Wu, Zhongchen Xiang, Qiaoli Yang, Chi Zhang, Wenxing Zhang, Xin Zhang, Dongning Zheng, Ruifeng Zheng, Jian-yong Zhou. 2024-04-16. Calibration of the Cryogenic Measurement System of a Resonant Haloscope Cavity. https://arxiv.org/abs/2404.10264