Modeling and Measurement of Lead Tip Heating in Implanted Wires with Loops

Abstract

Purpose: To theoretically and experimentally study implant lead tip heating caused by radiofrequency (RF) power deposition in different wire configurations that contain loop(s). Methods: Maximum temperature rise caused by RF heating was measured at 1.5T on 20 insulated, capped wires with various loop and straight segment configurations. The experimental results were compared with predictions from the previously reported simple exponential and the adapted transmission line models, as well as with a long-wavelength approximation. Results: Both models effectively predicted the trends in lead tip temperature rise for all the wire configurations, with the adapted transmission line model showing superior accuracy. For superior/inferior (S/I)-oriented wires, increasing the number of loops decreased the overall heating. However, when wires were oriented right/left (R/L) where the x-component of the electric field is negligible, additional loops increased the overall heating. Conclusion: The simple exponential and the adapted transmission line models previously developed for, and tested on, straight wires require no additional terms or further modification to account for RF heating in a variety of loop configurations. These results extend the usefulness of the models to manage implanted device lead tip heating and provide theoretical insight regarding the role of loops and electrical lengths in managing RF safety of implanted devices.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lydia J Bardwell Speltz, Seung-Kyun Lee, Yunhong Shu, Matt A Bernstein. 2024-12-19. Modeling and Measurement of Lead Tip Heating in Implanted Wires with Loops. https://arxiv.org/abs/2412.15120

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Prediction of biological radiation effects based on ionization clusters (nanodosimetry)

This article reviews approaches that link the formation of ionization clusters in nanometric volumes to radiobiological effectiveness. The corresponding models were developed as the field of nanodosimetry developed. Some address early biological radiation effects, such as DNA damage, while most aim to predict cell survival or inactivation. The models also differ in the nanodosimetric quantities considered, with many based on the probability distribution of ionization cluster formation in a single target. Some models account for the synergistic effects of pairs of ionization clusters formed in different targets. Several models feature macroscopic aggregation frameworks based on particle fluence, which are proposed for use in radiotherapy treatment planning, particularly in ion-beam radiotherapy. The models are presented here using harmonized terminology and notation for nanodosimetric quantities. An extension of the conceptual framework of nanodosimetry is also discussed. This extension transitions from a target-centered description to a track-centered description. It also introduces nanodosimetry-based analogs of dosimetric concepts, such as dose and linear energy transfer. This paper traces and summarizes the historical development of nanodosimetry-based biological effect models and discusses conceptual aspects of the models to reveal their underlying assumptions and the extent to which they are mechanistic or merely elucidate correlations. Eventually, an attempt is made to identify the key open questions in this field that still need to be addressed.

physics.med-ph

Contextual Cellular Growth (ConCeG) of neural cells for realistic grey matter tissue generation for diffusion MRI simulations

Accurate interpretation of diffusion magnetic resonance imaging (dMRI) signals in grey matter (GM) remains challenging due to the complex, heterogeneous, and densely packed cellular environment. Numerical phantoms provide a controlled framework for investigating the relationship between microstructure and diffusion signals, yet existing approaches often lack the morphological realism and multi-cellular organisation required to faithfully represent GM tissue. In this work, we introduce Contextual Cellular Growth (ConCeG), a generative framework for creating individual cells or constructing dense, three-dimensional, multi-cellular GM substrates informed by real neuronal and glial morphologies. The method combines topological neuron synthesis with a spatially constrained growth network, allowing for the controlled generation of heterogeneous cellular environments with realistic intra- and extracellular compartments. Synthetic cells are generated using morphological and topological characteristics derived from biological reconstructions. We validate the framework through comparisons of structural features with real cellular data, demonstrating strong agreement in branch order, length, angle, and tortuosity distributions. Power spectrum analysis further shows that both intracellular compartments reproduce the spatial correlations observed in biological tissue. Together, these results show ConCeG provides a biologically grounded framework for generating grey matter substrates suitable for large scale diffusion MRI simulation.

physics.med-ph

Magnetic Field of Firing Neuron in Humans: Measurable by Quantum Sensing MRI?

Firing neurons generate action potentials that propagate along axons to transmit signals supporting cognitive functions. These electrical currents generate magnetic fields, yet direct detection of these neuronal magnetic fields by MRI remains elusive. This Mini Review investigates why this goal has been proven difficult to achieve and whether an emerging approach, quantum sensing MRI, can overcome the challenge.

physics.med-ph