Search arXivSearch

arXiv · 2512.19120

Design, simulation and performance of the resistive-anode PICOSEC Micromegas detector

Abstract

The PICOSEC Micromegas detector is a Micro-Pattern Gaseous Detector concept developed to achieve tens of picosecond timing resolution for charged particle detection by combining a Cherenkov radiator with a two-stage Micromegas amplification structure. To improve operational robustness, a resistive anode has been implemented using a DLC layer deposited on a Kapton substrate. While this design enhances detector stability, the resistive layer may influence rate capability, signal formation, and detector capacitance, altering timing performance. This work presents a comprehensive study of a resistive design, including an analytical model and finite-element simulations to quantify rate-dependent gain reduction due to ohmic voltage drop on the resistive layer. An analytical solution for the voltage across a finite-size resistive layer is derived, and a numerical model is developed to evaluate gain suppression under intense particle fluxes. The impact of the resistive layer on signal formation is investigated using time-dependent weighting fields and the Garfield++ simulation framework. The contribution of signal components induced by the resistive layer is quantified, and preservation of the signal leading edge is found for surface resistivities above 100 kohm per square. Single-channel resistive-anode prototypes were designed, constructed, and experimentally characterized. Laboratory measurements using single photoelectrons and power spectral density analysis show the predicted reduction in signal amplitude while preserving the leading edge. Muon beam tests with CsI and DLC photocathodes demonstrate a time resolution of 11.5 ps for CsI, comparable to 11.9 ps for the metallic-anode device, showing the suitability of the resistive design for precision timing applications.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Djunes Janssens, Antonija Utrobicic, Marinko Kovacic, Marta Lisowska, Eraldo Oliveri, Florian Brunbauer, Karl Floethner, Hans Muller, Rui De Oliveira, Giorgio Orlandini, Leszek Ropelewski, Lucian Scharenberg, Thomas Schneider, Miranda van Stenis, Rob Veenhof, Thomas Papaevangelou, Roy Aleksan, Stephane Aune, Thomas Gustavsson, Alexandra Kallitsopoulou, Philippe Legou, Michal Pomorski, Emmanuel Scorsone, Lukas Sohl, Mariam Kebbiri, Spyros Tzamarias, Dimos Sampsonidis, Yannis Angelis, Ioannis Maniatis, Ioannis Karakoulias, Jonathan Bortfeldt, Ilaria Vai, Matteo Brunoldi, Davide Fiorina, Paolo Vitulo, Jaydeep Datta, Nathan Shankman, Kondo Gnanvo, Klaus Dehmelt, Brian Kross, Jack McKisson, Akash Pandey, George Fanourakis, Michele Gallinaro, Francisco Garcia, Yi Zhou, Jianbei Liu, Yue Meng, Xu Wang, Zhiyong Zhang, Michael Lupberger, Yorgos Tsipolitis, Maja Mičetić, Krešimir Salamon, Sebastian White. 2025-12-22. Design, simulation and performance of the resistive-anode PICOSEC Micromegas detector. https://arxiv.org/abs/2512.19120

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

TCAD + Allpi$\text{x}^2$ Simulation study of MALTA2, a Depleted Monolithic Active Pixel Sensor for future tracking

In this work, a hybrid simulation framework combining TCAD and Allpi$\text{x}^2$ is presented to investigate the sensor properties of MALTA2, a depleted monolithic active pixel sensor designed for future tracking. The study starts from 3D modeling and transient simulations in TCAD, with generic doping profiles and simple well structures. The resulting doping profiles and electric field are extracted and fed into Allpi$\text{x}^2$ for high-statistics Monte Carlo simulations in both DUT-only and full-telescope mode. Simulations reveal a strong dependence of sensor performance, specifically the detection efficiency and cluster size, on the doping concentration of the N-type blanket at the sensor surface. The doping concentration is then optimized by comparing simulations with measurement data. The active depth of the depleted region of the MALTA2 sensor is estimated in both simulations and measurements using a grazing angle method, in which the sensor is positioned at various inclinations relative to the beam, covering angles from 0 to 60 degrees. Excellent agreement on active depth is obtained with the optimal doping concentration, showing a deviation of 2\% from the measured value at a threshold of 450\,$\text{e}^-$. Consequently, the framework offers a generic toolkit for sensor studies without requiring proprietary information.

physics.ins-det

A compact fibre-coupled active beamstop for fast time-resolved monitoring of synchrotron X-ray beams

We have developed a compact fibre-coupled active beamstop for simultaneous interception and time-resolved monitoring of the direct X-ray beam in synchrotron diffraction experiments. The complete beamstop tip has a maximum transverse dimension of 1 mm and incorporates a 700 $μ$m-wide Ce:YAG scintillator. The scintillator converts the intercepted X-rays into visible light, which is transported through a multimode optical fibre to a silicon photodiode at the bottom of the blade that is connected to a dedicated fast amplifier electronics. Tests with a pulsed laser yielded a system decay time of 69.4 ns, consistent with reported luminescence decay times for Ce:YAG. Measurements at the P14.EH2 end station of PETRA III demonstrated the detection of individual bunch signals separated by 192 ns in the 40-bunch operating mode. The active beamstop also provided an in-situ timing signal for synchronising an experimental sequence with the approximately 190 $μ$s transmission window of a mechanical X-ray chopper. The device therefore combines a compact obstruction of the direct beam with fast monitoring of X-ray timing and flux during diffraction data acquisition.

physics.ins-det

Charge-Carrier transport simulations in diamond detectors with electric-field-dependent mobility and charge-collection-distance-based trapping

Diamond detectors are attractive for operation in harsh radiation environments because they combine radiation tolerance, fast signal formation, and low leakage current. Realistic detector-response simulations require an accurate description of charge-carrier mobility and trapping, which determine both signal amplitude and timing. In this work, we extend \allpix{}, a modular end-to-end detector simulation framework, with diamond-specific transport models. The implementation includes field-dependent mobility parameterizations for electrons and holes and an effective trapping model based on the charge collection distance (CCD), providing a detector-level interface to material quality and radiation-damage measurements. The mobility description is validated in the negligible-trapping limit using single-crystalline CVD diamond by comparing simulated drift velocities and transient-current signals with published reference data. For polycrystalline CVD diamond, the CCD-based trapping model is evaluated using experimentally measured CCD values and compared with laboratory transient-current-technique waveforms. The simulations reproduce the measured drift-velocity behavior in scCVD and the reduced charge collection and degraded transient response observed in pcCVD. The presented implementation enables detector-level studies of charge collection, pulse formation, and timing performance in diamond sensors using experimentally accessible transport and trapping parameters, and provides a practical framework for simulation-driven detector development and radiation-damage studies.

physics.ins-det